Points for Attention to Operation
. The BSFL- 03 Fluorospectrophotometer is suitably used in a laboratory room to perform
analytical measurements. If the instrument needs to be used at a worksite, the working
environment of the worksite should accord with the demands of the laboratory environment
basically.
2. If the user needs to move the instrument to other working site for analytical measurements,
please use original packing materials on moving.
3. Be sure to turn on∕off the power of the instrument according to the operating procedure.
When starting up the instrument, please turn on the power source of the xenon lamp first.
Then turn on the power source of the spectrophotometer main unit after the xenon lamp is lit.
When shutting down the instrument, please turn off the power source of the
spectrophotometer first and then the power source of the xenon lamp. Otherwise, the
instrument may not work normally. After the xenon lamp is lit, it is necessary to wait for 30
minutes over till the lamplight becomes stable. After the power source of the xenon lamp is
turned off, please wait for 60 seconds, and then turn on the lamp if the lamp needs to be used
again.
4. When the power source of the xenon lamp is turned on but the lamp fails to be lit and a
squeaking sound of high frequency or a cracking sound of sparking is produced continuously,
please turn off the power source of xenon lamp immediately and wait for several seconds,
then turn on the lamp again. Because the life of the xenon lamp is closely related to the
number of on-off of the lamp power source, the lamp should avoid being turned on/off
repeatedly under a high voltage and the unnecessary number of on/off of the xenon lamp
power source should be cut down as far as possible.
5. When the instrument is working, the surface temperature of the heat dissipation opening at
the top of the left back side of the instrument (xenon lamp position) becomes very high. Be
sure not to touch the surface of the heat dissipation opening with hands in order not to be
burnt. The air around the heat dissipation opening should be kept circulating.
6. After each startup of the instrument, please first confirm that the heat dissipation fan at the
side of the instrument runs normally in order to ensure normal work of the instrument. If the
user finds anything wrong with the heat dissipation fan, the instrument should be closed
down for an overhaul.
7. In order to protect the photomultiplier when PMT level is higher (>6), do not let high light