7
2-7 Auto Print
The auto print function allows the user to automatically send readings to a printer.
1. Press the setup key and use the •or –keys until “AUTO PRT” is displayed, then press the
yes key.
2. Use the •or –keys to toggle between “AUTO OFF” and “AUTO ON”. When the desired
printing function is displayed, press the yes key.
3. Press the meas key to exit the setup menu.
2-8 Auto Log
The auto log function allows the user to automatically save up to 100 test results in the
photometer’s memory. The procedure for using the auto log function is found in Chapter 4,
Section 4-2 Logging Data Automatically.
2-9 Self-Test
The self-test function puts the photometer into self-diagnostic mode. It is not recommended as
a routine test, but rather as a troubleshooting tool to be used if the unit gives an error message,
produces a suspect test result or in any way malfunctions. If an instrument fails any portion of
the self-test, contact CHEMetrics Technical Services Department.
1. Before initiating the self-test, the sample cell adapter must be removed and the light shield
must be in place.
2. Press the setup key and use the •or –keys until “SELFTEST” is displayed, then press the
yes key.
3. The instrument will display “KBD TEST”. Then it will prompt the user to press each of the
keys beginning with the 7 key.
4. When the test is complete, the instrument should display “UNIT OK”.
5. Press any key to view the complete LCD display capabilities. Compare the instrument display
to the display graphic in Chapter 1, Section 1-3 LCD Display.
6. Press any key to exit the self-test.
7. Press the meas key to exit the setup menu.
2-10 Upload
The upload function allows the user to update the V-2000 to the most current method program
revision. The V-2000 is updated by logging on to the CHEMetrics website, www.chemetrics.com
and clicking on the V-2000 Upload New Methods link. To ensure that the most current method
calibrations are being used, it is recommended that every V-2000 user check the CHEMetrics
website frequently. For more information, see Chapter 5, Section 5-3 Uploading Method
Revisions.