IMT Analytics CITREX H3 Manuale utente

User Manual
CITREX H3

Table of Contents
1 Introduction 5
2 Intended use 6
3 Safety instructions 7
3.1 Representation of hazards, cautions and notes 7
3.2 Personnel 7
3.3 Responsibility and guarantee 7
3.4 Service life 7
4 Symbolexplanation 8
5 Start-up 9
5.1 Power supply 10
5.2 Mechanical connectors 11
5.2.1 Flow channel 11
5.2.2 Oxygen sensor (option) 12
5.2.3 Installing the oxygen sensor 12
5.3 Electrical interfaces 14
5.4 Change CITREX battery 15
6 Operation 16
6.1 Switching the device on/off 16
6.2 Screen lock 16
6.3 Dim screen 16
6.4 User controls 17
6.5 Settings 18
6.5.1 Info display 18
6.5.2 Battery indicator 18
6.5.3 Ethernet interface 18
6.5.4 Trigger 19
6.5.5 Gas standard 19
6.5.6 Gas type 20
6.5.7 Gas humidity 20
6.5.8 Oxygen calibration (optional) 20
6.6 Numerical readings 21
6.7 Filter 21
6.8 Change parameters and units 22
7 Calibration 23
7.1 Zero point 23
7.2 Oxygen (O2) calibration 23
7.2.1 Calibration with air 24
7.2.2 Calibration with oxygen and air 24
8 Connecting the device 25
8.1 Generalmeasurement setup 25
8.2 Measurement setup for checking ventilators 26

9 Configuration tool 27
9.1 PC minimum requirements 27
9.2 Web server 27
9.2.1 Default 27
9.2.2 Configured 28
9.2.3 DHCP 28
9.2.4 Enable oxygen option 28
10 Servicing and care 29
10.1 Preventive cleaning and servicing operations 29
10.1.1 During operation 29
10.1.2 Every 4 weeks 29
10.1.3 Every 12 months 29
11 Accessories andspare parts 30
11.1 Accessories table 30
12 Disposal 31
13 Directives andapprovals 32
14 Specifications 33
14.1 Measurement parameter 33
14.2 Gas type 36
14.3 Power supply 36
14.4 Battery operation 36
15 Appendix 37
15.1 Principle of flow measurement 37
15.2 Trigger 37
15.2.1 Flow trigger 38
15.2.2 Pressure trigger 38
15.2.3 Base flow 38
15.2.4 Delay 38
15.3 Measurement parameters and units 39
15.4 Gas standards for flow and volume readings 40
15.5 Conversion factors 41
15.6 List of tables 42
15.7 List of figures 42
15.8 Index 43

Introduction
5
1
IMT Analytics AG
CITREX H3 was developed in order to measure flow and various pressures and
thus calculate various ventilation parameters. CITREX H3is a compact, mobile and
easy-to-operate measuring instrument. The device was developed in order to cali-
brate gas flows and pressures by simple means. With the oxygen option it is possible
to measure oxygen concentrations.
The descriptions and instructions in this manual refer to the product CITREX H3.
In this User Manual the unit "sL/min" is based on ambient conditions of 0°C and
1013.25mbar in accordance with DIN 1343.
This documentation applies to the followingversions:
CITREXH3 software: 4.4.000
CITREXH3 hardware: 4.0
In the case of older or newer versions there may be discrepancies in relation to this
User Manual.
Subject to technical modifications without notice.
Toavoid possible injuries,please read all the safety instructions be-
fore you use the product.
Thedevice is notintended for useoutside a building.
1 Introduction

Intended use
6
2
2
IMT Analytics AG
This product is intended for testing and calibration purposes on medical devices
orsystems thatgenerate gas flows or gas pressures. That includes ventilators and
anaesthetic equipment. The user of the device has received training on how to use
medical equipment and can perform repairs, maintenance and servicing on medical
devices. CITREX H3 can be used in hospitals, in clinics, at device manufacturers or
at independent service companies that perform repairs or servicing operations on
medical devices.
CITREX H3is intended for use in a laboratory environment. It may only be used out-
side the nursing sector. It must not be used directly on patients or devices that are
connected to patients. The measuring instrument CITREX H3is intended for over-
the-counter sale.
With CITREXH3 you have the solution for measurements in the following areas:
• Flow
• Pressure
• Volume
• Ambient pressure
• Temperature
In addition, various ventilation parameters can be measured:
• Ventilation rate
• Time
• Ratio
• Tidal volume
• Minute volume
• Peak flow
• Pressure
• Trigger
CITREX H3 is a measuring instrument for checking and calibrating
ventilators and anaesthetic equipment. It must not be used for patient
monitoring. During patient treatment by the ventilator it is not allowed
to connect to CITREX H3.
It is not allowed to measure liquids with CITREXH3.
2 Intended use

7
Safety instructions
3
IMT Analytics AG
Please read all the safety instructions carefully beforeyou use CITREX H3.
This User Manual uses the representation below to specifically draw attention to
residual risksduring intended use andemphasise important technical requirements.
Information and/orinstructions andprohibitions to preventdamage of any kind,as
well asuseful tipsand informationfor handling the device, will be indicated by the
following icon:
Work on and with CITREX H3may only be performed by persons
whohave undergone appropriate technical training andhave the
necessary experience.
Themanufacturer accepts no responsibilityor guarantee and willexempt itself fromli-
ability claims accordinglyif theoperatoror any third parties:
• Fail to use the device in accordance with its intended use.
• Disregard the specifications.
• Tamper with the device in any way (conversions, modifications or the like).
• Operate the device with accessories that are not listed in theassociatedsets
of product documentation.
Although the devicemeets high quality and safety standards andit
has beenconstructed and tested according to the current state
of the art,it is not possible to rule outthe risk of injuries withseri-
ousconsequencesif the device is usedinnon-compliance with the
intended use (improperly) or is misused.
Therefore please read through this User Manual carefully and keep
this documentation in a readily accessible place close to your device.
Themaximum service life of the devicehas been specified as 10(ten)years,provided
it is handled properly in accordance with this User Manual.
3 Safety instructions
3.1 Representation of haz-
ards, cautions and notes
3.2 Personnel
3.3 Responsibility and
guarantee
3.4 Service life

Symbol explanation
8
4
IMT Analytics AG
Thesymbols listed belowmayappear on the packaging material,on the device rating
plate and in the User Manual of the CITREX H3 measuring instrument.
USB interface
SN BBXXXX Serial number
Ethernet interface
On/Off button
SD card
Fragile contents
Keep dry
Read the User Manual
The device must not be disposed of in household waste
The device is CE approved
Caution: observe the safety instructions in the User Manual
Reusable packaging
Manufacturer's specification and date of manufacture
Keep away from heat
Temperature range for storage and transport
CSA monogram with C/US indicator
Direct current
Table 1: Symbol explanation
4 Symbolexplanation

9
Start-up
5
IMT Analytics AG
CITREXH3
Power supply plug with country-specific adapters
USB cable
MicroSD card
Dust filter RT019
Network cable
Table 2: Scope of delivery
5 Start-up

Start-up
10
5
IMT Analytics AG
CITREX H3can be operatedfrom the mains or from theintegrated battery.
Power can be supplied via the USB port (Mini-B) on the top of CITREXH3. Use the
power supply unit included to charge the battery or operate the device via the USB
port. You will find more information about power supply in the section "Electrical
interfaces".
During the charging process a green battery symbol is lit on the front.
Please connect the power supply unit included to a voltage of 100VAC to 240VAC
with a frequency of 50Hz to 60Hz.
USB port
Figure 1: Power supply
Before switching on, make sure theoperating voltage ofthe power
supply unitagrees with thelocalmains voltage. You will find this
information on the rating plate on the back of the power supply unit.
When operating CITREX H3via the USB port only use the original
power supply unit included!
Thedevice indicates visually and audiblywhen the battery has to be
charged. Please do not store the battery in a depleted state.
Caution: depletion can damage the battery beyond repair!