Kimo DBM610 Manuale utente

DBM610
AIRFLOW METER


Table of contents
1. Introduction....................................................................................................................................................................... 4
1.1. Key points................................................................................................................................................................. 4
1.2. Technical features..................................................................................................................................................... 4
1.3. General technical features........................................................................................................................................ 5
2. Presentation of airflow meter............................................................................................................................................ 6
2.1. ase.......................................................................................................................................................................... 6
2.2. Grid........................................................................................................................................................................... 7
2.3. Electronic housing..................................................................................................................................................... 7
Features:......................................................................................................................................................................... 7
2.3.1. Airflow meter mode........................................................................................................................................... 8
2.3.2. Micromanometer mode..................................................................................................................................... 8
2.3.3. Power supply of electronic housing................................................................................................................... 8
2.4. Hoods........................................................................................................................................................................ 9
3. Mounting the airflow meter................................................................................................................................................ 9
3.1. Mounting of frame..................................................................................................................................................... 9
3.2. Setting of the cloth.................................................................................................................................................. 10
3.3. Setting of rods......................................................................................................................................................... 11
4. Using D M610 in airflow meter mode............................................................................................................................. 12
4.1. Starting up............................................................................................................................................................... 12
4.2. Configure the airflow meter..................................................................................................................................... 12
4.2.1. Set the correction coefficient........................................................................................................................... 12
4.2.2. Set measurement units................................................................................................................................... 12
4.2.3. Set a damping................................................................................................................................................. 12
4.2.4. Activate or deactivate standardised airflow..................................................................................................... 12
4.3. Perform averages.................................................................................................................................................... 13
4.3.1. Perform an automatic average........................................................................................................................ 13
4.3.2. Perform an aggregate of automatic averages................................................................................................. 13
5. Using D M610 in micromanometer mode...................................................................................................................... 14
5.1. Starting up............................................................................................................................................................... 14
5.2. Set the micromanometer......................................................................................................................................... 14
5.2.1. Select the airflow sensor................................................................................................................................. 14
5.2.2. Set the surface................................................................................................................................................ 14
5.2.3. Set measurement units................................................................................................................................... 15
5.2.4. Set a damping................................................................................................................................................. 15
5.2.5. Activate or deactivate the solenoid valve........................................................................................................ 15
5.2.6. Set the thermocouple type.............................................................................................................................. 15
5.2.7. Set the compensation temperature................................................................................................................. 16
5.2.8. Activate or deactivate the standardised airflow............................................................................................... 16
5.3. Activate or deactivate air velocity and airflow.......................................................................................................... 16
5.4. Use the Matrix grid.................................................................................................................................................. 16
5.5. Perform averages.................................................................................................................................................... 16
5.5.1. Automatic average.......................................................................................................................................... 16
5.5.2. Point/point average......................................................................................................................................... 16
5.5.3. Point/point automatic average......................................................................................................................... 17
5.6. Perform an autozero............................................................................................................................................... 17
5.7. Hold the measurement............................................................................................................................................ 17
6. Manage dataset recordings............................................................................................................................................. 18
6.1. Get to recorded datasets......................................................................................................................................... 18
6.2. Delete recorded datasets........................................................................................................................................ 18
7. Set the device................................................................................................................................................................. 19
7.1. Set date and time.................................................................................................................................................... 19
7.2. Set language........................................................................................................................................................... 19
7.3. Set automatic shut-off............................................................................................................................................. 19
7.4. Set brightness......................................................................................................................................................... 19
7.5. Set contrast............................................................................................................................................................. 19
7.6. Activate or deactivate beep..................................................................................................................................... 19
8. Information about the device........................................................................................................................................... 20
8.1. Identification............................................................................................................................................................ 20
8.2. Calibration............................................................................................................................................................... 20
8.3. After sales service................................................................................................................................................... 20

1. Introduction
DBM610 airflow meter allows control and balancing of airflow in air conditioning system. Thanks to the interchangeable
hoods, the airflow can be used on any type of grids or diffusers in air supply or exhaust.
Easy to handle, lightweight, it allows accurate, sturdy and fast measurements.
t is supplied in standard with a 610 x 610 mm hood. Four other dimensions are available as options:
•1020 x 1020 mm
•720 x 720 mm
•720 x 1320 mm
•420 x 1520 mm
These hoods are airtight and have a transparent part on each side, allowing the user to see through the vent and
ensure that the hood is in a good position. The rods, made of fibreglass, provide a sturdiness of hoods.
The measurement grid, attached to the base, allows the measurement through 48 points all over the surface of the
grid. Measurement is performed using a differential pressure sensor, calibrated and compensated in atmospheric
pressure and temperature.
The measuring unit of the airflow meter, removable, can be used as a micromanometer
Real addition to airflow measurement, this instrument allows, by connecting a Pitot tube, to measure airflow velocity in air
duct or, by connecting two silicone tubes, to check filter fouling of air station. t is also possible to connect the matrix grid
designed for air velocity measurements on any types of large celling vent or diffuser. This grid calculates the average air
velocity from 24 points to guarantee a reliable measurement.
1.1. Key points
•Measuring range from 40 to 3500 m3/h
•Quick and simple hood changes
•Software for data processing on computer (datalogger 10)
•Sturdy, lightweight and easy to handle
•Removable measuring unit
•Interchangeable hoods
1.2. Technical features
•Airflow:
Measuring range: from 40 to +3500 m3/h
Accuracy: 3% of measuring ±10 m3/h
Resolution: 1 m3/h
•Air velocity:
Measuring range: from 0.2 to 10 m/s
Accuracy: 3% of measuring ±0.04 m/s
Resolution: 0.01 m/s up to 3 m/s and 0.1 m/s beyond
•Temperature (on the base)
Measuring range: from 0 to 50°C
Accuracy: 2% of measuring ±0.1°C
Resolution: 0.1°C
•Thermocouple Temperature (on the micromanometer)
Thermocouple type: K, J, T, S
Measuring range: K: from -200 to 1300°C
J: from -100 to 750°C
T: from -200 àto400°C
S: from 0 to 1760°C
4Introduction

Accuracy: K, J, T: from -200 to 0 °C: ±0.4°C ±0.3% of reading
from 0 to 1300 °C: ±0.4°C
S: ±0.6°C
Resolution: 0.1°C
•Pressure:
Measuring range: from -2500 to +2500 Pa
Accuracy: ±0.2% of measuring ±2 Pa
Resolution: from 0.001 to 0.1 Pa
Tolerated overpressure: 34473 Pa
1.3. General technical features
Display Large graphic back-lighted LCD / Sizes: 86 x 51 mm
Connectors ABS connectors, Ø 7 x 4 mm
Storage capacity 200 datasets of 100 points
Housing Shock-proof made of ABS, P54 protection
Keypad 12 keys
Conformity Electromagnetical compatibility (NF EN 61326-1 guideline)
Power supply 4 alkaline batteries LR6
Rechargeable battery Li-ion 3.7 V 4400 mhA (optional)
Environment Neutral gases
Operating temperature From 0 to +60 °C
Storage temperature From -20 to +70 °C
Auto-shut-off Adjustable from 0 to 120 min
Weight 3600 g
Languages French, English, German, Finnish, Dutch, talian, Norwegian, Portuguese, Slovak,
Swedish, Chinese
Introduction 5

2. Presentation of airflow meter
2.1. Base
•Outside of the base:
•nside the base:
6Presentation of airflow meter
Carrying handle
Protective cover Location for the
electronic housing
Integrated ergonomic handle
Button of point/point
measurement start
Electronic card
Silicone tube for pressure
routing
Receipt cup hood rods Measurement grid
Temperature probe
Two pressure connectors

2.2. Grid
The measurement grid is fixed to the base in 6 points and is made up of the following elements:
•two chambers (for total pressure and for static pressure)
•2 pressure connectors
The measurement grid allows to measure differential pressure. t takes into account atmospheric pressure and temperature
compensation and works automatically in supply or exhaust.
2.3. Electronic housing
Features:
•Shock-proof made of ABS
•P54 protection
•Large graphic back-lighted LCD (86 x 51 mm)
•Auto shut-off adjustable from 0 to 120 min
•Manual or automatic autozero
•Removable housing for micromanometer function
•4 magnets at the back allow the mounting on the base (airflow meter mode)
•2 pressure plugs allow to set the housing on the base and the connection of silicone tube for the measurement with
a Pitot tube
•1 thermocouple plug on the top of the housing (micromanometer mode)
Presentation of airflow meter 7
2 pressure
connectors
Validation button
Front side
On/Off button
Save button
Average button
Autozero button
Hold button
Menu button
Escape button
Low battery signal
Back side
Pressure plugs
Magnet
Thermocouple
plug
Battery trap

2.3.1. Airflow meter mode
n airflow meter mode, the device has following functions:
•Simultaneous display of airflow, pressure and temperature
•Automatic airflow direction (supply or exhaust)
•Automatic average and point by point average
•HOLD function (measurement is fixed)
•USB interface for PC operation
•Languages choice
2.3.2. Micromanometer mode
n micromanometer mode, the device has following functions:
•Measurement of air velocity with Pitot tube S, L or Debimo blades
•Measurement of thermocouple temperature
•Automatic average and point by point average
•HOLD function (measurement is fixed)
•USB interface for PC operation
•Languages choice
2.3.3. Power supply of electronic housing
The airflow meter is supplied in standard with 4 batteries type AA LR6 1.5V. Rechargeable batteries and Li- on battery are
available as options.
To replace batteries (rechargeable or not):
➢Remove the batteries cover then used batteries.
➢Disconnect the batteries support.
➢Replace batteries then reconnect the support.
➢Replace the batteries cover.
To put the Li-Ion battery (optional):
➢Remove the batteries cover then the batteries support.
➢Connect the battery.
➢Replace the batteries cover.
To load the device:
•f the device is on Li- on battery: connect the main adapter/charger on the side of the device
•f the device is on rechargeable batteries: remove batteries load them an external charger
t is also possible to load the electronic housing with the main adapter/charger supplied with Li- on battery.
8Presentation of airflow meter
Respect the connections (see below)
Connection for LR6 batteries
support (rechargeable or not)
Connection for
Li-Ion battery
Back view Side view
Connection for
main
adapter/charger
USB connection

2.4. Hoods
3. Mounting the airflow meter
3.1. Mounting of frame
To mount the different optional frames of airflow meter, just fit together the elements of the frame.
Stickers numbered from 1 to 8 are stuck on each end of the elements:
Example with 720 x 720 mm frame:
Mounting the airflow meter 9
Transparent mica allowing a good
visibility and making easy the position
of the hood
Elastic to hold the hood on
the base
Frame with a foam to ensure a
good airtight with ceiling
Frame rods
Step 1
Cloth of the hood
1122
3
3
4
4
5 566
7
7
8
8

3.2. etting of the cloth
Before set the cloth on the frame, frame must be mounted.
➢First, insert a corner of the frame inside a corner of the cloth.
➢nsert the opposite corner of the frame in the corresponding corner of the cloth.
The colour mark must be in the middle of the frame side. (1).
➢Press on the corresponding side of the cloth to put it in the rail of the frame.
➢Proceed the same way for the rest of the cloth.
When the cloth is fixed on the frame:
➢Place the cloth on the base.
The mark on the cloth and the Velcro tape are opposed to the carrying handle of the base.
➢Put the elastic of the cloth around the DBM base.
➢Tighten the Velcro tape of the cloth on the base.
➢f necessary, turn the cloth to match the mark of the cloth with the mark on the base.
10 Mounting the airflow meter
Zoom on the back of the
base with tightened Velcro
tape
Elastic to be set
around the base
Mark on the cloth Mark on the
base
Cloth on the base
1
Altri manuali per DBM610
1
Indice
Altri manuali Kimo Strumento di misura

Kimo
Kimo DS300 Manuale utente

Kimo
Kimo CT 100 0 Manuale utente

Kimo
Kimo SDA Manuale utente

Kimo
Kimo CT 110 Manuale utente

Kimo
Kimo KIGAZ 110 Manuale utente

Kimo
Kimo VT 200 Manuale utente

Kimo
Kimo MP130 Manuale utente

Kimo
Kimo LV 101 Manuale utente

Kimo
Kimo AQ 110 Manuale utente

Kimo
Kimo Luxmeter LX 100 Manuale utente

Kimo
Kimo DB300/1 Manuale utente

Kimo
Kimo CT 100 Series Manuale utente

Kimo
Kimo MP 200 P Manuale utente

Kimo
Kimo VT110 Manuale utente

Kimo
Kimo MP 202 Manuale

Kimo
Kimo CO 110 Manuale di istruzioni

Kimo
Kimo Kigaz-150 Manuale utente

Kimo
Kimo SDA Manuale utente

Kimo
Kimo SDA Manuale di istruzioni

Kimo
Kimo Kigaz 300 CLA Manuale utente