Linak JUMBO Care Manuale utente

Page 1 of 12
LINAK.COM/MEDLINE-CARELINE
Service readout and Service Data Tool 2
with JUMBO Care
User manual

Page 2 of 12
Contents
Preface ....................................................................................................................................... 3
Service intervals............................................................................................................................... 4
What is the service interval pre-set to?.......................................................................................... 4
How do I see on the control box that it is time for service?............................................................ 4
When does the service time starts counting?................................................................................. 4
How to change to another service interval?................................................................................... 4
Read out service data on a JUMBO Care with display ............................................................... 5
Read out service data on a laptop................................................................................................. 6
Equipment needed to read out service data................................................................................... 6
How to connect the equipment and get started............................................................................. 6
What to read out on the laptop .................................................................................................... 7
How to conclude on the service information.................................................................................. 8
How to exchange an actuator ........................................................................................................ 9
Resetting of service interval after service has been carried out....................................................... 9
Notice ........................................................................................................................................... 10
Addresses ..................................................................................................................................... 12

Page 3 of 12
Preface
We are delighted that you have chosen a product from LINAK.
Your new JUMBO Care control box has a microprocessor inside making it possible to
read out service data via the Service Data Tool and via the display (if available on your
control box). In this manual you can read about how to use and get the full benefit of
the service function in your JUMBO Care.
If no drivers are installed on your laptop (if you never have used service data tool on
your laptop before), please contact your local sales representative to get access to the
IB300001 user manual where this procedure is explained.
If you experience any problems with your new JUMBO Care you are always welcome to
contact your local LINAK representative who will be able to help you.
LINAK A/S

Page 4 of 12
Service Intervals
What is the service interval pre-set to?
A standard JUMBO Care control box will be pre-set with a service interval of:
12 months / 8000 cycles, whichever comes first.
How do I see on the control box that it is time for service?
The notice about service need depends upon the JUMBO Care version:
• No indicators on the front cover: one single beep telling that it is time for service
• Diodes on the front cover: The service diode will light up and one single beep will tell that
it is time for service
• Display: The display will show the service symbol and one single beep will tell that it is time
for service.
When does the service time starts counting?
The service time will start to count down from the day the control box is produced.
How to change to another service interval?
The service interval can be changed via a special LINAK handset (Item number HB8646V2010
+ 71). To change the service interval you simply plug-in the handset in the
JUMBO Care and press the button. You can choose from a number of months
or cycles between services. You can also choose ‘NO months’ or ‘NO cycles´.
E.g. Push M = 24, Push 10000.
The JUMBO Care is now set to indicate service need every 24 month or at
10000 cycles whichever comes first.
After having set a different service interval the control box will provide an
audio sound (as a receipt).

Page 5 of 12
Read out service data on a JUMBO Care with display
When you have a JUMBO Care with display it is possible
to have some basic service data on the display. To get this
information on the display press the “lifting arm up” button
on your LINAK hand control (short press ½ second).
The information that appears on the display is;
Total cycles done by the lifting actuator (channel 1)
Total work done by the lifting actuator (channel 1)
Total number of overloads (channel 1)
Days since last service/days between services
HB70HB80

Page 6 of 12
Read out service data on a laptop
Equipment needed to read out service data
When reading out service information on a laptop you need:
• Service data tool 2 version 1.3.4 or newer version installed on the laptop.
The software will be provided from your local LINAK sales representative.
• An OpenBus programming and data read out box (item number IB300001)
• A service readout cable (LINAK item number 0964478)
• One USB A-B cable
How to connect the equipment and get started
A) First of all make sure that service data tool drivers and the service data tool 2 software
(version 1.3.4 or higher) is installed on your laptop. If Service Data Tool drivers and the Service
Data Tool software are not installed please see the SDT2 user manual.
B) Connect the equipment as shown below.
JUMBO Care
When connection is observed the red LED (“Supply8VMissing”) in the current laptop window changes to green.
USB A-B
cable
Cable number:
0964478
C) Wake the control box (by pressing a button on the hand control or a button on the CB for
operation of lifting/lowering).
D) Enter the Start menu on your laptop and choose the LINAK programme to open the
Service Data Tool
The service read-out cable (item number 0964478) have to be connected via the HB port.
It is also possible to use a T-cable (item number 0015412) for connection of Hand control and Service Data Tool at the
same time.

Page 7 of 12
Ensure the JUMBO Care view is initiated by pressing the menu shown.
Go to the Diagnosis menu page 1
Days between services:
Can be changed via a special OpenBus HB
Days since last service:
Can be reset to 0 via pressing lifting arm up and lifting arm
down at the same time for 5 sec.
Cycles between service:
Can be changed via a special OpenBus HB
Cycles since last service:
Can be reset to 0 via pressing lifting arm up and lifting arm
down at the same time for 5 sec.
Total cycles:
Total work CH1:
Total work CH2:
Total overloads:
Can only be reset by exchanging the actuator through service
data tool statistics.
Battery status:
Figures 1-4 indicates the current status of the battery
Go to the Diagnosis menu page 2
What to read out on a laptop
When pressing a hand control button:
Input signals will give a receipt in the service data tool.
Note:
Only HB40 like hand controls will give signals.
OpenBusTM hand controls (HB70/HB80) or HD80 hand controls
do not give this function.

Page 8 of 12
How to conclude on the service information
Total cycles and total work:
Please contact the lifter manufacturer in order to decide when it is appropriate to consider
exchanging the actuator.
Overload:
If overload has occurred it is recommended to consider stronger lifting equipment with higher
working load for the particular patients/institution.
How to exchange an actuator:
When an actuator needs to be exchanged this has to be registered in the Service Data Tool via a
laptop.
Go to the Statistics menu.
Press the “replace” button to reset and mount a new actuator.
The action is confirmed with a shortly illustrated text at the buttom of the Service Data Tool screen:
“Service command CLEARED”
Activations In / Out:
No. of activations in in- or outwards direction
Endstops In / Out:
No. of times the actuator has reach endstop in in- or outwards direction
Overloads In / Out:
No. of times the actuator was overloaded, in- or outwards direction
Acc. work [A*s] In / Out:
Accumulated amount of work carried out, in- or outwards direction
[Amperage, A; time in seconds]
Acc. duration [s] In / Out:
Accumulated amount of time spent, in- or outwards direction [time in
seconds]
Replace Act. 1 / Act. 2:
When an actuator needs to be exchanged this has to be registered in the
Service Data Tool.
Replace actuator 1 (channel 1) or actuator 2 (channel 2) via button press to
reset the JUMBO Care - then mount a new actuator.

Page 9 of 12
Resetting of service interval after service has been carried out.
Resetting of service is done by pressing 2
buttons (lifting arm up and lifting arm down
on the hand control at the same time for 5
seconds. (Buttons that need to be pressed
might differ depending on hand control type).
After pressing the buttons for 5 seconds you
will receive an audio signal indicating that the
timer is reset. The timer will reset the diodes/
clear the display for service symbol, and start
counting a new service periode.
HB80 HB70
Resetting of service interval

Page 10 of 12
Notice:
• If the control box is exchanged the actuator data will be lost. The data is stored in the control box,
not the actuator.
• Please note that when you disconnect the service data tool from JUMBO Care it will take 1 hour
before the control box will power down. If you want to avoid the battery to consume energy the
emergency stop button can be activated.
• When a lifting actuator or a control box is exchanged it is neccessary to set the current cut off
value again via the learn mode function.
Indice
Altri manuali Linak Unità di controllo
Manuali Unità di controllo popolari di altre marche

Festo
Festo Compact Performance CP-FB6-E Manuale elenco delle parti

Elo TouchSystems
Elo TouchSystems DMS-SA19P-EXTME Manuale utente

JS Automation
JS Automation MPC3034A Manuale utente

JAUDT
JAUDT SW GII 6406 Series Guida rapida

Spektrum
Spektrum Air Module System Manuale utente

BOC Edwards
BOC Edwards Q Series Manuale utente