4. OPERATINGINSTRUCTION
4.1 AC + DC Voltage Measurement
WARNING
Maximum inputis600V.Do notattempttotakeany
voltagemeasurementthatexceedstheselimits.
Exceeding theselimitscould causeelectricalshock
anddamagetotheclampmeter.
1. Set the rotary switchto the “ V ” position.
2. Insert the test leads in to the input jack. (Black to COM
and Red to V)
3. Connect the test leads in PARALLEL to the measured
circuit.
4. The meter will automatically switch to ACV or DCV
display.
5. The meter willautomaticallyselect theappropriaterange.
6. Read the voltage and frequency values displayed on
theLCD. NOTE
Thesensitivityforvoltagefrequencymeasurementis1.2V
and the frequency range is 40 - 1KHz. If the frequency is
less than 40Hz the LCD may show Hz.
10. “ V mA ”Jack :
Connect red test lead for Capacitance, Resistance,
Diode, Countinuity and Micro ampere measurement
asa positiveterminal.
11. 3 Phase Adapter :
Insert plug in type 3 phase adapter in terminals
“COM”and“V”for3phase3wirepowermeasurement.
3. PARTS&CONTROLS
3.1 Description of Parts & Controls
2
5
4
10 8 9
3
7
6
1
11
Fig.
1
1. Transformerjaws(+marktofacesource):
TopickuptheACcurrentflowingthroughtheconductor.
2. Jawopeningtrigger
3. DataHold(MX/MN)button:
a. DataHoldMode:
Pressit once to hold the measured value and store the
value inmemory.Pressagaintoreleaseholdfunction.
(exceptcapacitancefunction)
b. MaximumandMinimumwithElapsedTimemode
0
1. Select desired ACA, ACV, DCV, C/ F or mA
function and perform the measurement, until the
measuredvalueshowsstablereadingontheLCD.
2. Press “ MX/MN ” button for 2 seconds to enter
recording with Elapsed Time mode, LCD will show
“ ® ” mark to lock the measurement range and
“ MX/MN ” mark to indicate the current measured
value with current elapsed time. The Auto Power
Offfunctionwillbeautomaticallycancelled.
3. Press “ MX/MN ” button one time, LCD will show
“ MX ” mark to indicate the recorded Maximum
valueandmaximumvalueoccurredElapsedTime.
4. Press “ MX/MN ” button one time, LCD will show
“MN ” marktoindicatetherecordedMinimum value
andminimumvalueoccurredElapsedTime.
5. Press “ MX/MN ” button one time, LCD will show
“ MX/MN ” mark to indicate the current value and
currentElapsedTime.
6. Step3,4,5canbecycled.
7. Press “ MX/MN ” button for 2 seconds to exit this
mode.
Note : The Elapsed Time intial setting is “ minute :
second ”, when current Elapsed Time is
over 60 minutes, the Elapsed Time will be
auto setting to “ hour : minute ”. The
maximum Elapsed Time length is 100
hours.
0
4. Functionselector:
Forselectionofdesiredfunction.
5. Rangebutton:
A. In ACA, ACV, DCV, mA, Capacitance and
Resistancefunction.
1. Press it once to enter the manual range select
mode,LCDwillshow®mark.
2. Pressagaintoselectthedesiredrange.
3. Press it 2 seconds to exit the manual range and
entertotheautorange mode, LCD ®markwill be
disappered.
B. InKW/HP/KVAfunction
a. Press RANGE button to select viewing the
KW/hp, KW/PF, KW/KVAR, KVA/uor V/A dual
display.
B1.Press HOLD button keep the measured valued
inmemory.
B2.Press RANGE button to select viewing the
KW/hp, KW/PF, KW/KVAR, KVA/uor V/A dual
display.
B3.PressHOLDbuttontoexitthismode.
C. InTEMPfunction 0 0
1. Press RANGE button to select desired C or F
temperatureunits.
6. Temperatureinputjack:
OnlytypeKthermocoupleinputisaccepted.
7. LCDDisplay:
4digital LCD with indicationsfor measurement values,
unit symbols, decimal point, polarity, over range and
lowbattery;etc.
8. COMJack:
Connect black test lead for voltage, Power,
Capacitance, Resistance, Diode, Countinuity and
Microamperemeasurement.
9. “V”Jack:
Connect red test lead for Voltage and Power
measurementasapositiveterminal.z