Met One Instruments E-FRM-DC Manuale utente

Met One Instruments, Inc.
1600 NW Washington Blvd.
Grants Pass, OR 97526
Telephone: (541) 471-7111
Facsimile: (541) 471-7116
www.metone.com
E-FRM-DC Reference Method Particulate Sampler Operation Manual - ©
Copyright 2013 Met One Instruments, Inc. All
Rights Reserved worldwide. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or
translated into any other language in any form without the express written permission of Met One Instruments, Inc.
E-FRM-DC
R
EFERENCE
M
ETHOD
P
ARTICULATE
S
AMPLER
O
PERATION
M
ANUAL
E-FRM-DC-9800,
R
EV
A

E-FRM-DC-9800 REV A Page 3
Table of Contents
1
INTRODUCTION 8
1.1
About This Manual ............................................................................................................ 8
1.2
Technical Service ............................................................................................................. 8
1.3
E-FRM-DC: Environmental Federal Reference Method Sampler ..................................... 9
1.3.1
E-FRM-DC U.S. EPA Configurations .............................................................................. 10
1.4
Model E-FRM-DC Specifications .................................................................................... 11
2
SITE SELECTION and POSITIONING CRITERIA 12
3
SETUP 14
3.1
Unpacking ....................................................................................................................... 14
3.2
Electrical Service ............................................................................................................ 15
3.3
Instrument Assembly and Deployment ........................................................................... 15
3.3.1
Assemble the E-FRM-DC Stand ..................................................................................... 16
3.3.2
Install E-FRM-DC sampler on top of Stand ..................................................................... 17
3.3.3
Installing the Ambient Temperature Sensor .................................................................... 17
3.3.4
Installing Inlet Down Tube and BX-802 PM
10
Inlet Head ................................................ 18
3.3.5
Configure the Internal Size Selection Hardware ............................................................. 19
3.3.5.1
Configuration for PM
2.5
Particulate .................................................................................. 19
3.3.5.2
Configuration for PM
10
Sampling .................................................................................... 20
3.3.6
Power Supply Installation ................................................................................................ 21
3.3.7
Power, Sensor, and Communication Connections .......................................................... 22
3.3.8
Dust Filter Removal and Installation ............................................................................... 24
3.4
Configurations for PM
2.5
or Other Particulate Cut-Points ................................................ 25
4
USER INTERFACE and MENU SYSTEM 26
4.1
The User Interface – Touchscreen Display Functions .................................................... 26
4.2
Main Operating Screens ................................................................................................. 26
4.3
Menu Hierarchy and Navigation ...................................................................................... 27
4.4
Operate Menu ................................................................................................................. 29
4.4.1
Pause/Stop Sample ........................................................................................................ 29
4.4.2
Event Manager ............................................................................................................... 30
4.4.3
Transfer Data .................................................................................................................. 30
4.4.4
USB Import ..................................................................................................................... 30
4.4.5
About .............................................................................................................................. 30
4.5
Test Menu ....................................................................................................................... 31
4.5.1
Temp Calibrate ............................................................................................................... 31
4.5.2
Pres Calibrate ................................................................................................................. 31
4.5.3
Flow Calibrate ................................................................................................................. 31
4.5.4
Leak Test ........................................................................................................................ 32
4.5.5
Parameters ..................................................................................................................... 32
4.5.6
Fan Test .......................................................................................................................... 32
4.6
Setup Menu .................................................................................................................... 33
4.6.1
Event Defaults ................................................................................................................ 33
4.6.2
Standard Temp ............................................................................................................... 34

Page 4 E-FRM-DC-9800 REV A
4.6.3
Clear Memory ................................................................................................................. 34
4.6.4
Set Clock ........................................................................................................................ 35
4.6.5
Unit ID ............................................................................................................................. 35
4.6.6
Password ........................................................................................................................ 35
4.6.7
Serial Port ....................................................................................................................... 35
4.6.8
Sound Volume ................................................................................................................ 36
4.6.9
Touch Calibrate .............................................................................................................. 36
4.7
Alarms Menu ................................................................................................................... 36
5
SCHEDULING SAMPLER EVENTS and SAMPLER OPERATION 37
5.1
Configuring New Sample Events Using the Event Manager ........................................... 37
5.2
Modifying Sample Events Using the Event Manager ...................................................... 37
5.3
Deleting Sample Events Using the Event Manager ........................................................ 38
5.4
Viewing Current and Previous Sample Events ............................................................... 38
5.5
Importing New Sample Events from a USB Flash Drive ................................................. 39
6
FLOW SYSTEM and FLOW CALIBRATIONS 40
6.1
Flow System Diagram ..................................................................................................... 40
6.2
Leak Tests and Flow Audits ............................................................................................ 40
6.3
Leak Tests ...................................................................................................................... 41
6.3.1
The External Leak Test ................................................................................................... 41
6.3.2
Internal Leak Test ........................................................................................................... 42
6.4
Temperature, Pressure, and Flow Audits........................................................................ 43
6.4.1
Auditing the Ambient Air Temperature Sensor ................................................................ 44
6.4.2
Auditing the Filter Temperature Sensor .......................................................................... 44
6.4.3
Auditing the Filter and Box Pressure Sensors ................................................................ 45
6.4.4
Auditing the Sampler Flow Rate ..................................................................................... 45
6.5
Multi-Point Flow Calibrations .......................................................................................... 47
7
MAINTENANCE AND TROUBLESHOOTING 49
7.1
Recommended Periodic Maintenance Table .................................................................. 49
7.2
Ambient and Filter Temperature Calibrations ................................................................. 49
7.3
Filter and Box Barometric Pressure Sensor Calibrations ................................................ 50
7.4
Error and Alarm Descriptions .......................................................................................... 51
7.5
Basic Problem and Cause/Solution Table ....................................................................... 52
7.6
Leak Correction .............................................................................................................. 52
7.6.1
Correcting an External Leak ........................................................................................... 53
7.6.2
Correcting an Internal Leak ............................................................................................. 53
8
DATA RETRIEVAL 54
8.1
Using the RS-232 or USB Cable Connections ................................................................ 54
8.1.1
User Communication ...................................................................................................... 54
8.1.2
Computer Communication .............................................................................................. 56
8.2
Using a USB Flash Drive ................................................................................................ 57
8.3
Using FSCommAQ Software .......................................................................................... 57
9
SPARE PARTS and ACCESSORIES 58

E-FRM-DC-9800 REV A Page 5
9.1
Consumables, Replacement Parts, and Accessories ..................................................... 58
Table of Figures
Figure 1-1 E-FRM-DC Air Sampler ..................................................................................................... 9
Figure 3-1 E-FRM-DC Front View with Door Closed and Opened .................................................... 15
Figure 3-2 E-FRM-DC Sampler Stand Components ......................................................................... 16
Figure 3-3 E-FRM-DC Sampler Stand Final Assembly ..................................................................... 16
Figure 3-4 Securing the E-FRM-DC Sampler to the Stand ............................................................... 17
Figure 3-5 Ambient Temperature Sensor Assembly Mounting ......................................................... 17
Figure 3-6 Installation of the PM
10
Inlet ............................................................................................. 18
Figure 3-7 Installation of the PM
2.5
Cyclonic Separator ..................................................................... 19
Figure 3-8 Installation of the PM
10
Pass-Through Adapter ............................................................... 20
Figure 3-9 Securing the Power Supply Case .................................................................................... 21
Figure 3-10 Bottom of E-FRM-DC Enclosure .................................................................................... 22
Figure 3-11 Cable Connections ........................................................................................................ 22
Figure 3-12 Cable Penetration Seal .................................................................................................. 22
Figure 3-13 Dust Filter Installation .................................................................................................... 24
Figure 4-1 The E-FRM-DC Main Operating Screen .......................................................................... 26
Figure 4-2 The E-FRM-DC Main Screen ........................................................................................... 26
Figure 4-3 Main Menu Drop Down Selections .................................................................................. 28
Figure 4-4 Setup Menu ..................................................................................................................... 28
Figure 4-5 Clock Setup Screen ......................................................................................................... 28
Figure 4-6 Visual Keypad for Numeric Entry ..................................................................................... 28
Figure 4-7 Keypad for Alphanumeric Entry ....................................................................................... 29
Figure 4-8 The Operate Menu .......................................................................................................... 29
Figure 4-9 Pause Sample Sequence ................................................................................................ 29
Figure 4-10 Resume Sample Sequence ........................................................................................... 30
Figure 4-11 The Event Manager Screen ........................................................................................... 30
Figure 4-12 The About Screen .......................................................................................................... 30
Figure 4-13 The Test Menu .............................................................................................................. 31
Figure 4-14 The Temp Calibrate Screen........................................................................................... 31
Figure 4-15 The Pres Calibrate Screen ............................................................................................ 31
Figure 4-16 The Flow Calibrate Screen ............................................................................................ 31
Figure 4-17 The Leak Test Screen ................................................................................................... 32
Figure 4-18 The Test Menu Parameters Screens ............................................................................. 32
Figure 4-19 The Fans Screen ........................................................................................................... 32
Figure 4-20 The Setup Menu Screen ................................................................................................ 33
Figure 4-21 The Event Defaults Screen ............................................................................................ 33
Figure 4-22 The Event Defaults Setting Screens .............................................................................. 33

Page 6 E-FRM-DC-9800 REV A
Figure 4-23 The Standard Temp Screen .......................................................................................... 34
Figure 4-24 The Clear Memory Screen............................................................................................. 34
Figure 4-25 The Clear Memory Confirmation Screen ....................................................................... 34
Figure 4-26 The Set Clock Screen .................................................................................................... 35
Figure 4-27 The Unit ID Screen ........................................................................................................ 35
Figure 4-28 The Password Screen ................................................................................................... 35
Figure 4-29 The Serial Port Screen .................................................................................................. 35
Figure 4-30 The Sound Volume Screen............................................................................................ 36
Figure 4-31 The Touch Calibrate Screen .......................................................................................... 36
Figure 4-32 The Alarms Menu .......................................................................................................... 36
Figure 5-1 The Event Add Screen .................................................................................................... 37
Figure 5-2 The Event Manager Screen ............................................................................................. 37
Figure 5-3 The Event Delete Screen ................................................................................................ 38
Figure 5-4 The Event Viewer History Screen .................................................................................... 38
Figure 5-5 The USB Import Sequence .............................................................................................. 39
Figure 6-1 E-FRM-DC Flow System Diagram ................................................................................... 40
Figure 6-2 External Leak Test Cassette Insertion Screen ................................................................. 41
Figure 6-3 External Leak Test Sequence .......................................................................................... 41
Figure 6-4 Leak Test Cassette Removal Screen .............................................................................. 42
Figure 6-5 Internal Leak Test Cassette Insertion Screen .................................................................. 42
Figure 6-6 Internal Leak Test Sequence ........................................................................................... 43
Figure 6-7 Ambient Temperature Audit with BX-307 ........................................................................ 44
Figure 6-8 Filter Temperature Sensor Location ................................................................................ 44
Figure 6-9 Filter Temperature Sensor Removed .............................................................................. 44
Figure 6-10 Flow Calibrate Cassette Insertion Screen ..................................................................... 46
Figure 6-11 Flow Calibration Flow Rate Selection Screen ................................................................ 46
Figure 6-12 Flow Calibration Cassette Removal Screen .................................................................. 46
Figure 6-13 Flow Calibrate Cassette Insertion Screen ..................................................................... 47
Figure 6-14 Flow Calibration Flow Rate Selection Screen ................................................................ 47
Figure 8-1 Data Transfer Interface Screens ...................................................................................... 57

E-FRM-DC-9800 REV A Page 7
Table of Tables
Table 4-1 Main Display Parameter Descriptions ............................................................................... 27
Table 4-2 E-FRM-DC Menu Hierarchy .............................................................................................. 27
Table 5-1 E-FRM Current Status Screen Descriptions ..................................................................... 39
Table 7-1 Recommended maintenance Schedule for the E-FRM-DC Sampler ................................ 49
Table 7-2 The E-FRM-DC Alarm Conditions ..................................................................................... 51
Table 8-1 Common Serial Command Set ......................................................................................... 55

Page 8 E-FRM-DC-9800 REV A
1 INTRODUCTION
1.1 About This Manual
This document is organized with the most important information toward the front of the manual. All
users should read and understand the sections on setup, operation, and field audits. Toward the
back are sections that provide in-depth information on subjects such as diagnostics and
accessories. These sections should be consulted as needed.
This manual is periodically revised for maximum accuracy and to incorporate new features or
updates. User feedback is welcome. Electronic versions of this manual are available upon request.
1.2 Technical Service
Should support still be required after consulting the printed documentation, contact one of the expert
Met One Instruments, Inc. Technical Service representatives during normal business hours of 7:00
a.m. to 4:00 p.m. Pacific Standard Time, Monday through Friday. In addition, technical information
and service bulletins are often posted on our website. Please contact us and obtain a Return
Authorization (RA) number before sending any equipment back to the factory. This allows us to
track and schedule service work and to expedite customer service.
Contact
Information:
Tel: + 541 471 7111
Fax: + 541 471 7115
Web: http:/www.metone.com
Email: [email protected]
Address: Met One Instruments, Inc.
1600 Washington Blvd
Grants Pass, Oregon
97526 U.S.A.
Please have the instrument serial number available when contacting the manufacturer. On most
models manufactured by Met One Instruments, it will be located on a silver product label on the unit,
and printed on the calibration certificate. The serial number will begin with a letter and be followed
by a unique five digit number such as U15915.

E-FRM-DC-9800 REV A Page 9
1.3 E-FRM-DC: Environmental Federal Reference Method Sampler
Figure 1-1 E-FRM-DC Air Sampler
The Met One Instruments, Inc. Model E-FRM-DC is a programmable single-event filter sampler
designed to meet federal specifications as a reference method for the determination of daily PM
2.5
or
PM
10
particulate concentrations in ambient air.
It employs a single pre-weighed 47 mm sample filter disc in an EPA-standardized cassette which
must be manually exchanged by the operator prior to each sample period. The instrument runs a
sample period based on an operator programmed start time and sample duration.
A vacuum pump, controlled to 16.67 liters-per-minute, draws ambient air through a series of
standardized inlets that provide the PM
2.5
or PM
10
cut points. The sample then passes through the
filter, where the airborne particulate matter is deposited.
The dirty filter is collected after the sample period and sent to a lab to be equilibrated and reweighed
per the standardized EPA gravimetric procedures. The resulting clean and dirty mass values for the
filter are then correlated to the volume of air sampled to determine the concentration of particulate in
micrograms-per-cubic-meter (µg/m3)
The E-FRM-DC can easily be configured for reference method PM
2.5
sampling using the
standardized BGI VSCC™ very-sharp-cut cyclone or the classic WINS Impactor.

Page 10 E-FRM-DC-9800 REV A
Alternatively, other sharp-cut cyclones can be installed for non-designated or indicative PM2.5 or
PM1 sampling.
If PM
10
monitoring is desired, the PM
2.5
cyclones can be omitted entirely, and a pass-through
adapter can be installed in its place.
1.3.1 E-FRM-DC U.S. EPA Configurations
The
E-FRM-DC
is US-EPA designated under the following designation numbers:
• RFPS-0315-221: PM
2.5
designation
• RFPS-0216-231: PM
10
designation
• RFPS-0316-232: PM
10-2.5
designation
US-EPA designated methods using the E-FRM-DC are modified occasionally to reflect hardware or
software improvements. These modifications do not impact previously designated configurations of
the E-FRM-DC but may provide the end user with a product upgrade path that will allow the monitor
to continue to be operated as a US-EPA designated method. For further details, please contact our
service department. Details concerning USEPA designated configurations of the E-FRM-DC may be
found on the US-EPA website.

E-FRM-DC-9800 REV A Page 11
1.4 Model E-FRM-DC Specifications
PARAMETER SPECIFICATION
Measurement Principle: Programmable-event filter sampler for determination of particulate matter
concentrations
Resulting Data Units: Micrograms per actual or standard cubic meter µg/m
3
. Calculated from lab filter
weights and actual or standard sampled volumes.
Design Specifications: Title 40 CFR, Part 50 Appendix L
U.S. EPA Designations: RFPS-0315-221, RFPS-0216-231, RFPS-0316-232
Main Inlet Type: EPA-pattern PM
10
size-selective louvered inlet standard (MOI part no. BX-802)
PM
2.5
Separator Type: Options: BX-808 Very Sharp Cut Cyclone or BX-804 WINS Impactor
PM
10
Sampling Config: Pass-through tube (used in place of cyclone) included for PM
10
sampling.
Other Configurations: Compatible with PM
2.5
or PM
1
SCC cyclones. TSP inlet available.
Filter Type: 47mm PTFE disc filters standard. Other filter material types available.
Filter Holder: Accommodates standard EPA-pattern white Delrin filter cassettes.
Filter Temp Control: Filter Temp within 5°C of ambient during sampling and standby. Logged in data.
Operating Temperature: -30 to +50°C Ambient
Operating Pressure: 400 to 800 mmHg, Ambient
Sample Events: 24 hour samples standard. Programmable start date/time and duration.
Flow Rate: 16.67 L/min (1 m
3
/hr.). With flow sensor and real-time actual flow control.
Flow Accuracy: ±2% (±.33 L/min)
Flow Records: Flow rate, temperature, pressure logged in 5 minute intervals.
Vacuum Pump: Single-head diaphragm 12 VDC pump
Upper Concentration
Limit: At least 200 µg/m
3
estimated for daily average.*
Data Collection: RS-232 port (9-pin). USB data port. USB flash drive port. Software supplied.
Memory Capacity: 24 sample events
Compatible Software: FSCommAQ
User Interface: 4.3” graphical touch screen
Enclosure Type: Active fan ventilation
Deployed Dimensions: 2.0 meters (78.75”) high, 65.5 cm (25.75”) wide, 64.1 cm (25.25”) deep
Footprint Dimensions 65.5 cm (25.75”) wide, 64.1 cm (25.25”) deep
Weight: Sampler 33 Ib (except PM
10
inlet head); Stand 10.5 Ib
Power: 100 to 240VAC, 33W nominal (2.75A)
* The approximate upper concentration limit is based on the sampler’s ability to maintain 16.67 L/min sample flow over the entire 24 hour period,
despite PTFE filter loading with typical urban PM
2.5
particulate. Actual concentration limits will vary depending on ambient conditions, altitude, and
particulate type.
Specifications may be subject to change without notice.
Indice
Altri manuali Met One Instruments Strumento di misura

Met One Instruments
Met One Instruments ES-412 Manuale utente

Met One Instruments
Met One Instruments GT-526 Manuale utente

Met One Instruments
Met One Instruments BX-597A Manuale utente

Met One Instruments
Met One Instruments BC 1054 Manuale utente

Met One Instruments
Met One Instruments 083F Manuale utente

Met One Instruments
Met One Instruments BX-807 Manuale utente

Met One Instruments
Met One Instruments MSO Manuale utente

Met One Instruments
Met One Instruments Swift 25.0 Manuale utente

Met One Instruments
Met One Instruments 076B Manuale utente

Met One Instruments
Met One Instruments 085A Manuale utente