
Doc# 100911 • REV B (May 2016) Page 6 of 8
4.4 Test Procedure
1. Before the dynamic test with the Dynamic Signal Checker the measuring loop sensivity has to be
determined, since deviaons from the specied sensivity inuence the accuracy and indicaon. Using
a stac calibrang unit like the METRIX 9600-001 and a calibrated voltmeter you can verify that probe,
cable and driver match the specicaon
(typically: 7.87 V/mm, 200mV/mil).
2. Place the Dynamic Signal Checker on a solid surface.
3. Screw the GAP adjusng ring + lock ring up to about 1 mm into the upper part of the housing.
4. Connect probe including extension cable with driver/transmier.
5. Connect the voltmeter with the driver/transmier. Set the voltmeter on “DC” in order to read out
the GAP voltage.
6. Carefully insert the probe or the probe holder with reverse-mount probe, into the GAP adjustment
ring and clamp it in the prism holder at a gap signal of about 10 VDC. Avoid contact with the vibrator in
order to not damage vibrator or probe p. Fine adjustment can be achieved by turning of the adjusng
ring; then x it with the lock ring.
The prism holder is suitable for probe diameters of 8 … 22 mm.
7. Connect the simulaon unit to the handheld control unit.
8. Aer pressing the ON/OFF (“ ”) buon once, the device performs a self-test (ashing of all 5 LEDs).
As soon as the “ON” status indicator lights up, the device is ready for operaon.
9. By further pressing of the “ ” buon one by one up to 4 stored default amplitude values (see 11.
for adjustment) can be selected. The amplitude can be read from the voltmeter as RMS voltage signal
[mV] and converted to [µm pp, mils pp] via calculaon formula (p. 4), conversion table (p. 5) or diagram
(p. 6).
10. If the vibraon indicated in the control room doesn’t correspond with the mV signal of the
voltmeter, this may be due to one of the following causes:
• The wiring of the measuring loop is not correct or even broken.
• The vibraon driver/transmier is defecve or calibrated with a wrong measuring range.
• The limit monitoring in the PLC downstream of the transmier is calibrated with
a wrong measuring range.
• The input card of the vibraon monitor is defecve or falsely adjusted, thus a wrong value
is transferred to the monitor.
• The measuring range of the vibraon indicaon is falsely adjusted.
11. In order to adjust individual default values (possible within the range of 0 … 250 µm, 0 to 10 mils)
following steps are necessary:
• Connect a calibrated voltmeter to the buered-out of the exisng driver/transmier or
monitor and set it to “VAC”.
• Read the corresponding mV value of the desired vibraon amplitude from the table (p. 4)
or diagram (p. 5).
• Adjust the vibraon amplitude according to the mV value by means of pushing the buons
“ ” resp. “ ”. You can choose between holding the buons (fast change of values) or
just tapping the buons (slow change of values for rather sensive adjustment).
• The adjusted value is to be compared with the operaon indicaons.
4.4 Test Procedure
• In order to save the individually adjusted value, as shortcut level for future adjustments,
select one of the 4 storage posions (green LEDs) with the buon “ ”. By pressing of
the buon “S” the stored value will be replaced by the new selected value. Note
that, depending on the baery status (14.) and the posion (2.) of the Dynamic
Signal Checker the actual amplitude corresponding to the stored values may slightly dier
over me. The correct seng can be updated anyme.
12. In order to switch the device o manually, the ON/OFF (“ ”) buon has to be pressed for > 2
seconds. If no operaon is performed over a period > 30 minutes, the device switches o automacally.
13. If the “ON” indicator blinks, the device has to be charged. Before damaging exhausve discharge
starts, the device is automacally switched o; 5 seconds ahead all LEDs start blinking.
14. In case of low baery the device can be powered using the USB charging cable.