MRC CD-4309 Manuale utente

CONDUCTIVITY METER
Model : CD-4309
Your purchase of this
CONDUCTIVITY METER
marks a step forward for you
into the field of
precision measurement.
Although this meter a
complex and delicate
instrument, its durable
structure will allow
many years of use if
proper operating
techniques are
developed. Please read
the following
instructions carefully
and always keep this
manual within easy
reach.
OPERATION MANUAL
3, Hagavish st. Israel 58817 Tel: 972 35595252, Fax: 972 3 5594529
MRC. 10.18

TABLE OF CONTENTS
1. FEATURES.........................................................................................
1
2. SPECIFICATIONS.........................................................................................
2
2-1 General Specifications................................................................
2
2-2 Electrical Specifications........................................................................
4
3. FRONT PANEL DESCRIPTION...........................................................................
6
4. FUNCTION SELECTION.......................................................................
7
5. CONDUCTIVITY/TDS MEASURING and
CALIBRATION PROCEDURE....................................................................
8
5-1 Conductivity measurement..........................................................
8
5-2 TDS ( PPM ) measurement..................................................................
9
5-3 Calibration................................................................................ 9
6. SALT MEASURING and calibration PROCEDURE............................
12
6-1 Salt measurement..........................................................................
12
6-2 Hardness measurement..........................................................................
12
6-3 Resistance measurement..........................................................................
13
6-4 Calibration............................................................................................
13
7. OTHER FUNCTION.....................................................................................
14
7-1 Data Hold.........................................................................................
14
7-2 Record ( Max./ Min. reading )....................................................................
14
7-3 LCD Backlight ON/OFF....................................................................
15
8. ADVANCED SETTING...............................................................................
16
9. POWER SUPPLY from DC ADAPTER.....................................................
19
10. BATTERY REPLACEMENT.............................................................
19
11. SYSTEM RESET.................................................................................
20
12. RS232 PC SERIAL INTERFACE.....................................................
20
13. OPTIONAL ACCESSORIES..........................................................................
22

1. FEATURES
* One meter for multi purpose operation : Conductivity,
TDS (Total dissolved solids), Salt , Hardness , Resistance measurement.
* Conductivity : 20 uS/ 200 uS/2 mS/20 mS/200 mS.
* Salt : 0 to 12 % salt ( % weight ).
* Hardness : 0 to 100,000 ppm .
* Resistance : 5 ohm to 99.99 M ohm .
* Conductivity measurement can select Temp. Coefficient
of measurement solution.
* ATC for the conductivity measurement.
* Separate probe, easy for operation of different measurement environment.
* LCD with green light backlight, easy reading.
* It can default auto power off or manual power off.
* Data hold, record max. and min. reading.
* Microcomputer circuit, high accuracy.
* Power by UM3/AA ( 1.5 V ) x 6 batteries or DC 9V adapter.
* RS232/USB PC COMPUTER interface.
* Wide applications: water conditioning, aquariums,beverage,
fish hatcheries, food processing, photography,laboratory, paper industry,
plating industry, quality control, school & college, water conditioning.
1

2. SPECIFICATIONS
2-1 General Specifications
Circuit Custom one-chip of microprocessor LSI
circuit.
Display LCD size : 52 mm x 38 mm
LCD with green backlight ( ON/OFF ).
Measurement * Conductivity ( uS, mS )
Function * TDS ( Total Dissolved Solids, PPM )
* Salt ( % Weight )
* Temperature ( ℃,℉)
Temperature Automatic from 0 to 60 ℃(32 - 140 ℉),
Compensation with temperature compensation factor
variable between 0 to 5.0% per C.
Conductivity Pt glass electrode for long life. Hight performance
Probe for the low conductivity ( pure water ) measurement.
Advanced * Set clock time (Year/Month/Date,Hour/Minute/ Second )
setting * Set sampling time
* Auto power OFF management
* Set beep Sound ON/OFF
* Set temperature unit to ℃or ℉
* Set CD temperature compensation factor
* Set probe CELL K Value
Data Hold Freeze the display reading.
Memory Recall Maximum & Minimum value.
2

Sampling Time Approx. 1 second.
of Display
Data Output RS 232/USB PC computer interface.
* Connect the optional RS232 cable
UPCB-02 will get the RS232 plug.
* Connect the optional USB cable
USB-01 will get the USB plug.
Operating Meter 0 to 50 ℃.
Temperature Probe 0 to 60 ℃.
Operating Less than 85% R.H.
Humidity
Power Supply
* AC/DC power adapter is optional.
Alkaline or heavy duty DC 1.5 V battery
( UM3, AA ) x 6 PCs, or equivalent.
* AC/DC power adapter is optional.
DC 9V adapter input. ( AC/DC power
adapter is optional ).
Power Current Normal operation ( w/o SD card save
data and LCD Backlight is OFF) :
Approx. DC 6 mA.
* AC/DC power adapter is optional.
If LCD backlight on, the power
consumption will increase approx.
12 mA.
Weight 489 g/1.08 LB.
Dimension Meter 177 x 68 x 45 mm
(7.0 x 2.7x 1.9 inch)
Probe Round,
22 mm Dia. x 120 mm length.
Accessories *
Instruction manual.....................................................
1 PC
Included * Conductivity/TDS/Salt pt glass probe,
CDPB-04...........................................................
1 PC
*
Hard carrying case ( CA-06 )..................................
1 PC
3

Optional * 1.413 mS Conductivity Standard
Accessories
Solution.......................................................
CD-14
AC to DC 9V adapter.
USB cable, USB-01.
RS232 cable, UPCB-02.
Data Acquisition software,SW-U801-WIN.
2-2 Electrical Specifications (23±5
℃
℃℃
℃
)
Conductivity ( uS, mS )
Range Measurement Resolution Accuracy
20 uS 0 to 20.00 uS 0.01 uS
200 uS 0 to 200.0 uS 0.1 uS
2 mS 0.2 to 2.000 mS 0.001 mS ±(2% F.S.+1d)
20 mS 2 to 20.00 mS 0.01 mS * F.S. -
200 mS 20 to 200.0 mS 0.1 mS full scale
* Temperature Compensation :
Automatic from 0 to 60 ℃( 32 - 140 ℉), with temperature
compensation factor variable between 0 to 5.0% per C.
* The accuracy is specified under measurement value ≦100 mS.
* mS - milli Simens * @ 23±5℃
TDS ( Total Dissolved Solids )
Range Measurement Resolution Accuracy
20 PPM 0 to 13.20 PPM 0.01 PPM
200 PPM 0 to 132 PPM 0.1 PPM
2,000 PPM 132 to 1,320 PPM 1 PPM ±(2% F.S.+1d)
20,000 PPM 1,320 to 13,200 PPM 10 PPM * F.S. -
200,000 PPM 13,200 to 132,000 PPM 100 PPM full scale
* Temperature Compensation :
Automatic from 0 to 60 ℃( 32 - 140 ℉), with temperature
compensation factor variable between 0 to 5.0% per ℃.
*The accuracy is specified under measurement value ≦66,000 PPM.
* PPM - parts per million * @ 23±5℃
4

Temperature
Function Measuring Range Resolution Accuracy
℃0 ℃to 60 ℃0.1 ℃±0.8 ℃
℉32 ℉to 140 ℉0.1 ℉±1.5 ℉
* @ 23±5℃
Salt
Measurement 0 to 12 % salt ( % weight ).
Range
Resolution 0.01 % salt.
Accuracy ± 0.5 % salt value
Hardness
Range Measurement Resolution Accuracy
10 PPM 0 to 10.00 PPM 0.01 PPM
100 PPM 0 to 100 PPM 0.1 PPM
1,000 PPM 100 to 1,000 PPM 1 PPM ±(2% F.S.+1d)
10,000 PPM 1,000 to 10,000 PPM 10 PPM * F.S. -
100,000 PPM 10,000 to 100,000 PPM 100 PPM full scale
Resistance
Range
Measurement Resolution Accuracy
automatic Range 5 ohm to 99.99 M ohm 1 ohm ±(2% F.S.+1d)
0.01 M ohm * F.S. -full scale
5

3. FRONT PANEL DESCRIPTION
Fig.1
3-1 Display
3-2 Power Button ( Backlight Button )
3-3 Hold Button
3-4 REC Button
3-5 Range Button ( ▲Button )
3-6 Function Button ( ▼Button )
3-7 Time Button (SET Button)
3-8 Enter Button ( Logger Button )
3-9 Stand
3-10 Battery Compartment/Cover
3-11 Battery Cover Screw
3-12 Tripod Fix Nut
3-13 Probe Socket
3-14 DC 9V Power Adapter Input Socket 3-18 Probe Handle
3-15 Reset Button 3-19 Sensing head
3-16 RS-232 Output Terminal 3-20 Probe Plug
6

4. FUNCTION SELECTION
1) Power ON the meter by pressing and holding the " Power
Button " ( 3-2, Fig. 1 )for at least 1.5 seconds .
*Pressing the " Power Button " ( 3-2, Fig. 1 )
continuously and > 1.5 seconds again will turn off the
meter.
2) The meter can select 5 kind Mode as :
a.
Conductivity measurement
b.
TDS measurement
c.
Salt measurement
d.
Hardness measurement
e.
R
esistance
measurement
pressing and holding the " FUNCTION Button " ( 3-6, Fig. 1 )
continuously , Display will show the following text in sequence :
C
D
Conductivity measurement
TD
S
TDS measurement
SAL
T
Salt measurement
HD
Hardness measurement
R
Resistance
measurement
Until the Display show the desired mode the meter will
execute this FUNCTION with default.
7

5. CONDUCTIVITY/TDS MEASURING
and CALIBRATION PROCEDURE
The meter default function are following :
* The display unit is set to conductivity ( uS, mS ).
* The temperature unit is set to ℃.
* Temp. compensation factor is set to 2.0% per C.
* Auto range.
* Auto power off disable.
* The sampling time of data logger function is
2 seconds.
Fig. 2
5-1 Conductivity measurement
1) Prepare the Conductivity Probe ( CDPB-04, standard
accessory, included ), install the " Probe Plug " ( 3-20,
Fig. 1) into the " Probe Socket " ( 3-13, Fig. 1 ).
2) Power on the meter by pressing and holding " Power Button "
( 3-2, Fig. 1 ) for at least 1.5 seconds .
Select the Meter's measurement FUNCTION to " CD "
( Conductivity measurement ), refer to chapter 4, page 7 .
3) Hold the " Probe Handle " ( 3-18, Fig. 1 ) by hand
and let the " Sensing head " ( 3-19, Fig. 1 )
immersed wholly into the measured solution. Shake
the probe to let the probe's internal air bubble drift
out from the sensing head.
Display will show the conductivity mS ( uS ) values.
at the same time the left bottom display will show the
Temp. value of the measured solution.
Manual range operation
The meter is default to be used for the auto range
mode. by pressing and holding the " Range Button " ( 3-5, Fig. 1 )
continuously in sequence will change the range from 20uS ,
200 uS, 2 mS , 20 mS, 200 mS and auto range.
8
Indice
Altri manuali MRC Strumento di misura

MRC
MRC UT-FLU5B Manuale utente

MRC
MRC SPECTRO-96 Manuale utente

MRC
MRC CD-4303 Manuale utente

MRC
MRC MS-7004SD Manuale utente

MRC
MRC DW-6090 Manuale utente

MRC
MRC LX-1148SD Manuale utente

MRC
MRC TM-8812 Manuale utente

MRC
MRC DT-2236 Manuale utente

MRC
MRC HE218B-EX Manuale utente

MRC
MRC FG-6020SD Manuale utente

MRC
MRC SP-82AM Manuale utente

MRC
MRC VIS-Series Manuale utente

MRC
MRC HE2 Series Manuale utente

MRC
MRC PC-6011SD Manuale utente

MRC
MRC PH-207 Manuale utente

MRC
MRC PM-1064SD Manuale utente

MRC
MRC RSW-923 Manuale utente

MRC
MRC EM-9300SD Manuale utente

MRC
MRC PMG-302 Manuale utente

MRC
MRC BCT-4308 Manuale utente