•Detection Method: Catalytic Oxidation + Double FID
•Measuring Range: 0~10/100/1000/10000 ppm (Automatic)
•Detection Limit: ≤50ppb (Calculated by Carbon)
•Zero Drift: ≤±1% Full Scale/8H
•Span Drift: ≤±1% Full Scale/8H
•Repeatability: ≤±0.1% Full Scale
•Indication Error: ≤±1% Full Scale
•Conversion Efficiency: ≥95%
•Warm-up Time ≤30min
•Data Resolution (T90): THC≤2s; NMHC≤2s
•Working Environment: temperature: -20 ℃~50 ℃; relative humidity: <95%; atmospheric
pressure: (80~106) kPa
•Power Supply: DC24V, AC (100-240±10%) V/(50±2%) Hz
•Dimensions: 370x260x397mm
•Weight: <12KG
Technical Data
Standard Configuration
Instrument main unit (including catalytic module, metal hydride cylinder), sampling tube
(including probe, filter device), Microsoft Surface GO tablet, backpack.
Lightweight & Integrated Design
•Highly integrated design to ensure the total weight within 12KG (including catalytic module,
metal hydride cylinder, battery, zero air generators, etc)
•Analyze THC & NMHC in one single unit.
•Equipped with a backpack and a handle, which can be easily transported and put into use by
one person.
Easy to Use
•Large-capacity battery (45Ah) power supply, which can run for a long time even without
utility power on site.
•Operated by a detachable touch screen tablet with Wi-Fi communication and a user-friendly
interface. Users can control the instrument and read test results within 50m (33ft) of the
site through the detachable tablet.
Features