Omicron Lab Bode 100 Manuale utente

Smart Measurement Solutions®
Bode 100 - Information
Detailed Functional & Calibration Check of
Bode 100
Table of Contents
1CHECK OF FACTORY CALIBRATION ..............................................................................................................2
1.1 TEST OF OUTPUT SIGNAL SHAPE ...................................................................................................................................2
1.2 ACCURACY OF OUTPUT FREQUENCY.............................................................................................................................4
1.3 ACCURACY OF OUTPUT LEVEL.......................................................................................................................................6
1.4 ACCURACY OF GAIN/PHASE FACTORY CALIBRATION....................................................................................................7
1.5 NOISE FLOOR /CHANNEL ISOLATION...........................................................................................................................10
1.6 ACCURACY OF IMPEDANCE CALIBRATION....................................................................................................................12
1.7 QUICK CHECK OF ATTENUATORS.................................................................................................................................14
1.8 CHECK OF ATTENUATOR CORRECTION........................................................................................................................16
1.9 QUICK CHECK OF AC COUPLING CAPACITORS AT INPUT CHANNELS 1&2.................................................................18
1.10 CHECK OF INTERNAL CONNECTIONS &OUTPUT IMPEDANCE....................................................................................19
2CALIBRATION SERVICE...................................................................................................................................20
By Bernhard Baumgartner & Tobias Schuster
© 2008-2020 by OMICRON Lab –V4.0
Visit www.omicron-lab.com for more information.
Contact support@omicron-lab.com for technical support.

Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 2 of 21
Smart Measurement Solutions
Smart Measurement Solutions®
1 Check of factory calibration
The following checks will help you to decide if it is required to send the Bode 100 to OMICRON Lab
for calibration.
Note: It is assumed that the user manual of the Bode 100 was read and understood prior the
measurements described in this document are executed. If you do not have a manual at
hand you can download it at https://www.omicron-lab.com/downloads/vector-network-
analysis/bode-100/.
1.1 Test of output signal shape
Measurement Setup:
Please connect the Source output of Bode 100 to an oscilloscope
Equipment Setup:
Please select the measurement type “Gain / Phase”, select the source settings as follows and start a
continuous measurement:
Frequency: 10 MHz - Fixed
Output level: 0 dBm
Figure 2: test signal shape - measurement type setting
Figure 1: test signal shape - connection setup

Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 3 of 21
Smart Measurement Solutions
Smart Measurement Solutions®
Figure 3: test signal shape - measurement settings
Figure 4: test signal shape - start continuous measurement
Expected Measurement result:
Figure 5: test signal shape - expected result
Waveform: sinusoidal
Vpeak-peak: ~ 1.26 Volts (more accurate measurement follows)
Frequency: ~ 10 MHz (more accurate measurement follows)
If the waveform looks nice, the source amplifiers are working properly.

Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 4 of 21
Smart Measurement Solutions
Smart Measurement Solutions®
1.2 Accuracy of output frequency
Measurement Setup:
Please connect the output of the Bode 100 to a frequency counter which is locked to a high stable
frequency normal (e.g. Rubidium, GPS etc.).
Figure 6: output frequency accuracy - connection setup

Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 5 of 21
Smart Measurement Solutions
Smart Measurement Solutions®
Equipment Setup:
Please select the measurement type “Gain / Phase”, select the source settings as follows and start a
continuous measurement:
Frequency: 10 MHz - Fixed
Output level: 0 dBm
Figure 7: output frequency accuracy - measurement type setting
Figure 8: output frequency accuracy - measurement setting
Figure 9: output frequency accuracy - start continuous measurement
Expected Measurement result:
Measured frequency: 10 MHz
Maximum tolerance for Bode 100 Revision 1:
± 15 ppm ≙± 150 Hz @ 10 MHz (< 1 year after adjustment)
± 25 ppm ≙± 250 Hz @ 10 MHz (< 3 years after adjustment)
Maximum tolerance for Bode 100 Revision 2:
± 2 ppm ≙± 20 Hz @ 10 MHz (< 1 year after adjustment)
± 4 ppm ≙± 40 Hz @ 10 MHz (< 3 years after adjustment)
If you measure a higher frequency deviation, please send the Bode 100 in for calibration / adjustment.

Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 6 of 21
Smart Measurement Solutions
Smart Measurement Solutions®
1.3 Accuracy of output level
Measurement Setup:
Please connect the Source output of Bode 100 to a calibrated 50 Ωpower meter.
Figure 10: output level accuracy - connection setup
Equipment Setup:
Please select the measurement type “Gain / Phase”, select the source settings as follows and start a
continuous measurement:
Frequency: 1 MHz –Fixed (or whatever frequency your power meter is made for)
Output level: 0 dBm / 13 dBm / -27 dBm
Figure 11: output level accuracy - measurement type setting
Figure 12: output level accuracy - measurement settings
Figure 13: output level accuracy - start continuous measurement

Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 7 of 21
Smart Measurement Solutions
Smart Measurement Solutions®
Expected Measurement result:
Measured output power: 0 dBm / 13 dBm / - 27 dBm
Measured output power tolerance @ 23°C ± 5°C:
max. ± 0.3 dB (1 Hz to < 1 MHz)
max. ± 0.6 dB (1 MHz to 40 MHz for Bode 100 R1)
max. ± 0.6 dB (1 MHz to 50 MHz for Bode 100 R2)
If you measure a higher power deviation, please send in the Bode 100 for calibration / adjustment.
1.4 Accuracy of Gain/Phase factory calibration
Measurement Setup:
Connect the Source of Bode 100 to the two inputs using the cables and the
T-BNC Adapter delivered with Bode 100.
Figure 14: gain/phase accuracy - connection setup
Equipment Setup:
Please select the measurement type “Gain / Phase” and apply the following settings in the hardware
setup:
Figure 15: gain/phase accuracy - gain/phase measurement
Figure 16: gain/phase accuracy - open hardware setup

Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 8 of 21
Smart Measurement Solutions
Smart Measurement Solutions®
Figure 17: gain/phase accuracy - hardware setup
Figure 18: gain/phase accuracy - measurement settings
Then start a single measurement.
Figure 19: gain/phase accuracy - start single sweep
For better visibility change the chart setup to “one axis per chart” as well as set the trace settings as
can be seen below:
Figure 20: gain/phase accuracy - chart setup

Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 9 of 21
Smart Measurement Solutions
Smart Measurement Solutions®
Figure 21: gain/phase accuracy - trace settings
Expected Measurement results:
Gain: 0 dB
Tolerance: max. ± 0.2 dB from 1 Hz to 50 MHz
Phase: 0°
Tolerance: max ± 3° from 1 Hz to 50 MHz
If you measure a higher gain/phase deviation please send in the Bode 100 for calibration /
adjustment.
The following measurement results are an example and can be slightly different from the
measurement with other devices.
Figure 22: gain/ phase accuracy –magnitude (dB) measurement

Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 10 of 21
Smart Measurement Solutions
Smart Measurement Solutions®
Figure 23: gain/phase accuracy - phase (°) measurement
1.5 Noise floor / Channel isolation
Measurement Setup:
Disconnect all BNC cables from Bode 100.
Do not connect anything to the front side BNC connectors!
Equipment Setup:
Please select the measurement type “transmission / reflection” and apply the following settings:
Figure 24: dynamic range / isolation - transmission / reflection measurement
Altri manuali per Bode 100
3
Indice
Altri manuali Omicron Lab Strumento di misura

Omicron Lab
Omicron Lab CT Analyzer Manuale utente

Omicron Lab
Omicron Lab OTMC 100 Manuale utente

Omicron Lab
Omicron Lab SPECTANO 100 Manuale utente

Omicron Lab
Omicron Lab SPECTANO 100 Manuale utente

Omicron Lab
Omicron Lab Bode 100 Manuale utente

Omicron Lab
Omicron Lab CP AL1 Manuale utente

Omicron Lab
Omicron Lab SPECTANO 100 Manuale utente