PDK PDR1000 Manuale utente

PDR1000
Digital Pressure Gauge
Users Manual
25

26
Table of Contents
1. General Introduction ……………………………………………………………………………………… 27
2. Abbreviations & Initials ………………………………………………………………………………… 28
3. Introduction ………………………………………………………………………………………………… 28
3.1 Specification …………………………………………………………………………………… 29
3.2 Front / Back panel …………………………………………………………………………… 30
4. Operation ………………………………………………………………………………………………… 32
4.1 Key Operation
…………………………………………………………………………………… 32
4.1.1 Power On …………………………………………………………………………… 32
4.1.2 PEAK …………………………………………………………………………………… 32
4.1.3 UNIT …………………………………………………………………………………… 32
4.1.4 ZERO …………………………………………………………………………………… 32
4.1.5 Back Light …………………………………………………………………………… 33
4.2 Menu/Function Select ……………………………………………………………………… 33
4.2.1 Alarm …………………………………………………………………………………… 33
4.2.2 Pressure Switch Test ……………………………………………………………… 35
4.2.3 Calibration …………………………………………………………………………… 38
4.2.4 Baud Rate …………………………………………………………………………… 39
4.2.5 RS232 Communication …………………………………………………………… 40
4.2.6 Auto Power Off …………………………………………………………………… 42
4.2.7 Data Logger ………………………………………………………………………… 43
4.2.8 Storage Data Clear ……………………………………………………………… 45
4.2.9 End …………………………………………………………………………………… 45
5. Maintenance Adjustments and Calibration ……………………………………………………… 46
5.1 Zero Adjustment ……………………………………………………………………………… 46
5.2 Pressure Calibration …………………………………………………………………………… 46
5.3 Analog Output Calibration ………………………………………………………………… 46
6. Troubleshooting ……………………………………………………………………………………………… 47
7. Appendix
…………………………………………………………………………………………………… 48
7.1. Warranty Statement
………………………………………………………………………… 48
7.2. Accessory
……………………………………………………………………………………… 48
7.3. Table of pressure unit conversion ……………………………………………………… 49

1. General Introduction
Manual Conventions
This manual is intended to provide the user with the basic information necessary to operate a
PDR1000 (Digital Pressure Gauge). It also includes a great deal of additional information provided
to allow you to optimize PDR1000 use and take full advantage of its many features and functions.
“Caution” is used in throughout the manual to identify conditions or actions that could
cause harm to the PDR1000 or user of the PDR1000.
“Note” is used throughout the manual to identify operating and applications advice
and additional explanations.
High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and
gases can be released unexpectedly and with extreme force. High pressure systems should be
assembled and operated only by authorized personnel who have been instructed in proper safety
practices.
27

2. Abbreviations & Initials
Units & Words Keys Units & Words Keys
kPa kilopascal a absolute
MPa megapascal g gauge
hPa hectopascal PEAK peak
mbar millibar Hi high
bar bar Lo low
psi pounds per square inch MAX maximum
mmH2O millimeter of water MIN minimum
mmHg millimeter of mercury Hold Hold
inH2O inch of water mA Milliampere
inHg inch of mercury Alarm
atm atmosphere V Voltage
kgf/cm2kilogram force per square
centimeter REM serial interface
communication standard
3. Introduction
The PDR1000 Series Reference Digital Pressure Gauges (the Product) are high accuracy digital
pressure test gauges. PDR1000 has best performance that can measure precision pressure from
-100 kPa to 500 MPa without a separate external power supply in anytime any place, including
on-site and laboratory. Also PDR1000 equipped solid and stylish aluminum outer casing for
durability.
PDK produced PDR1000 with PDK’s unique pressure calibration and production technology
that has best performance in its class. PDR1000 is digital pressure gauge for precise pressure
measurement and test with Built-in various features.
Simply use buttons on front panel to operate and select various functions and external power
input section, analog output port (1 to 5 VDC), and communication port (RS232) and batteries
change socket on rear panel.
Combine pressure comparator or hand pump for using pressure calibrator in field and calibration
lab for simple pressure measurement and calibration provides an excellent solution.
28

3.1 Specification
Pressure Range
Accuracy
Over Pressure limit
Bust Pressure
Pressure Unit
Workable
Temperature Range
Storage
Temperature Range
Temperature
Compensation Range
RS232 Communication
Power Requirement
Media of Use
Display
Display Speed
Analog Output
Pressure Port
Weight
Size
Data Logging
-100 kPa to 150 MPa 0 to 250 MPa 0 to 500 MPa
Gauge : ±0.025% F.S
Absolute / Compound : ±0.05% F.S ±0.05% F.S ±0.1% F.S
** Included Nonlinearity, Hysteresis, Repeatability, errors for
-10 ~ 50 ℃ Temperature range
150 % of full scale
kPa, MPa, kg/cm2, psi, mbar, bar, inHg, inH2O, mmH2O, mmHg, atm
-20 ~ 70 ℃
-30 ~ 80 ℃
-10 ~ 50 ℃
Provide Commands in manual,
Available to use private communication cable (Option)
AA Alkaline battery 3ea,
Hours of battery use approximately 1000 hours
External power supply(Option, 12~24 VDC),
Gas & Liquid (Only Gas available below 100 kPa)
5 Digits, Backlight On/Off, Auto Off
3 times / Second
(10 times automatic change if Peak function use)
1 - 5 VDC (Must use external power supply), 0.1% Accuracy
(Option : 4-20mA , 0-5VDC , 0-10VDC)
1/4" NPT M, 1/4" PF (G), 1/4" PT (R), 9/16" UNF Cone Threaded
(HF4/AF250C, 100 MPa / 150 MPa)
530 g
Dia. 110 (W) mm x 38 (D) mm x 150 (H) mm Included Pressure Port
1 time / 1, 3, 5, 30, 60 second, Store Max 3000 data
Refer to pressure type and range in catalog.
29

3.2 Front / Back Panel
3.2.1 On / Off Key
PDR1000 Power On/Off key
3.2.2 PEAK/MENU Mode key
3.2.3 UNITS / Analog out / Down Key
3.2.4 ZERO / Up Key
PEAK
MENU
UNITS ZERO
Photo 1. PDR1000 Front / Rear Panels Photo
Figure 1. Buttons
Peak shows maximum and minimum pressure value.
Menu makes selection and set for Alarm, Switch, CAL, Baud Rate, Auto Power Off, and Data
Logger. Saved peak value will be clear when PDR1000 turns off.
Change the pressure units and On/Off for Analog output.
In configure mode, push Down key to move backward through the menus.
(Analog output function is only available when external power supply used)
Zeroing pressure offset value.
Pressure Switch On/Off, Peak value will be clear when Peak mode.
In configure mode, push Up key to move forward through the menus.
30

3.2.5 Back Light / Enter Key
Backlight On/Off
Use Enter key to set up the Menu.
[ D-Sub 9 Pin Map ]
Figure 2. RS-232 Port
Table 1. Pin outs
5
6789
4 3 2 1
Pin No Description Note
1GND
2RX RS232
3TX
4Analog Out+ Analog Output
5GND
6Ext. Switch In
7Alarm Out
8Ext. Power + External Power
9Ext. Power -
31

4. Operation
4.1 Key Operation
4.1.1 Power On
: Push POWER key on front panel to power on.
Screen at no pressurize or Zero set.
4.1.2 PEAK / MENU
: Peak shows maximum and minimum pressurized value on PDR1000.
Push PEAK/MENU key indicates Hi for maximum pressurized value.
Push PEAK/MENU key again, indicates Lo for minimum pressurized value.
Push PEAK/MENU key again, return to the current pressure value.
4.1.3 UNITS
: This function can change engineering units on PDR1000.
When push UNIT key engineering units will change as following steps.
MPa →mbar →bar →inH2O →mmH2O →inHg →mmHg →kgf/cm² →atm
→psi →kPa →MPa
To set Analog Out function, push UNIT key for 2 seconds.
(It only operates when external power supply connected)
4.1.4 ZERO
: This function is zeroing the PDR1000.
ZERO
PEAK
MENU
UNITS
For Gauge mode push ZERO key for 2 seconds under normal atmosphere condition. For the
32
psi
%
g
0

4.1.5 Back Light / Enter
: Push to turn on the backlight. Push again to turn off the backlight.
4.2 Menu/Function Select
4.2.1 Alarm
Alarm function must need external power supply. (Figure 3 reference)
Figure 3. Alarm Setup flowchart
Absolute mode, make vacuum state (under 1 × 10-2 torr) then ZERO key for 2 seconds to zeros
the display. Zero key can clear the Peak Hi Low values when Peak mode. Also it can select the
set up when Pressure Switch mode.
Push Enter key for 2 seconds to start or stop for Data Logger function.
To set Data Logger function, turn on the Data Logger function on configuration menu.
Push PEAK/MENU key for 2 seconds to go to setup and configuration menus.
Push power key to go to End Menu or Push PEAK/MENU key to set other function settings after
Menu function sets.
Analog out, Alarm : These function must need external power supply to operate.
33
ZERO
psi
%
g
0
psi
%
g
0
Alarm
Lo
psi
%
g
0
Alarm
Hi
psi
%
g
0
Alarm
psi
%
g
0
Alarm
psi
%
g
0
Alarm
psi
%
g
0
Alarm
%
0
psi g
PEAK
MENU

First screen of Alarm function page. Press UP key to turn Alarm On.
Alarm On screen on above. Press Enter key to save Alarm On status then Alarm Low value
setup screen shows with blink.
Alarm Low value setup screen on above. Digits on above screen is percentage value of PDR1000
Full Scale. Push ZERO key and UNITS key to change percentage value.
After push Enter key to save Alarm Low value, Alarm Hi value setup screen shows with blink.
Alarm Hi value setup screen on above. Digits on above screen is percentage value of PDR1000
Full Scale. Push ZERO key and UNITS key to change percentage value.
After push Enter key to save Alarm Hi value, Electric Contact Alarm Signal setup screen shows
with blinking.
34
psi
%
g
0
Alarm
Lo
psi
%
g
0
Alarm
Hi
psi
%
g
0
Alarm
psi
%
g
0
Alarm