9
MRM-10
4. Measurement
Check that the item to be measured is voltage free.
1. Before turning the equipment on, connect the test leads to the item to be measured and to the front panel
terminals.
2. Using the test current selector, choose the range and the current to be used.
3. Choose the power supply to use. If you are going to use the battery, advance to the following point. If you
are going to use the mains supply, plug the power cord into the connector.
4. Switch the equipment On using the On/Off.
5. The PRESS START message will appear showing that measurement can be started. Press the Start key.
Depending on the position of the test current control the LOW CURRENT message will turn up.
6. Turning the test current control clockwise increases the current. Adjust this control until obtaining the
desired value or until the current indicator (bar graph) indicates 100%. The bar graph will show the test
current value as a percentage of the nominal value selected by using the test current selector.
7. The lowest current for measuring is 10% of the nominal value. It is important to consider that the
measurement errors increase while test current decreases. The equipment accuracy is specified for test
currents higher than 80%.
8. The display will show the resistance value measured and the corresponding unit (Ω [ohms], mΩ [milli-
ohms] or µΩ [micro-ohms]).
9. The value can be retained in the display by pressing the hold key. Pressing this key again releases the
value.
10. Press the Stop key in order to finish the measurement. Do not turn Off the equipment without pressing the
Stop key.
11. Finally, after finishing measurements, turn the equipment Off using the On/Off switch.
CAUTION: Do not connect or disconnect the test leads
during the measurement.