5
8681 and 8682 are capable of measuring
with Automatic Temperature Compensation.
"ATC" shows under "CAL" at the left corner
of the screen. To select the temp. unit
o o
(C or F) preferred, turn off the meter first.
AUTOMATIC TEMPERATURE
COMPENSATION (ATC): ONLY
FOR 8681/8682
While the meter is off, press " " and
o o
" " at the same time until the C or F
appeares on the LCD. Press " " to s elect
the preferred unit and then press " " to
save. " " will appear on the LCD for one
second and then back to normal (Fig.C&D)
CC
Fig. C Fig. D
1. Power on the meter
2. Place the electrode into a buffer solution
(4,7or10), pH7 should be calibrated first
and then 4 or 10 pH for better accuracy
3. Press " " (8681/8682) or " "
(8680) to enter pH calibration mode.
The text "CA"( or CAL, only for 8682)
will appear on the LCD for one second
(Fig.E) and then text "CAL" and solution
pH value (4,7 or 10) will be displayed on
the LCD. (Fig. F&G)
Calibration is necessary and should be
done regularly, recommend everyday if
the meter is used often. The unique
calibration design of the meter features
automatic buffer recognition to avoid errors.
CALIBRATION MODE (CAL)