RadiSense® Product Manual
This product manual refers to the RadiSense® series.
RadiSense® Models: RSS2010E - By Raditeq
RadiSense® LASER card: LPS2001B - By Raditeq
Publish Date: 15/08/2022
All trademarks used in this manual are the property of their respective owners
Supplier Information
Raditeq B.V.
Vijzelmolenlaan 3
3447 GX, Woerden
The Netherlands
Tel.: +31 (0)348 200 100
Internet: www.raditeq.com
Email: [email protected]
Carefully read the content of this manual before operating the product and make sure all the safety
instructions are strictly followed.
For your convenience, a Quick Start Guide has been added to this product. This Quick Start Guide
contains the basic start-up steps and main safety warnings.
Please keep the Quick Start Guide (and this regular manual) close at hand when you operate your new
Raditeq product(s).
Please contact your local reseller if you have any questions.
RadiSense® Product Manual
RSS2010E
Page 2 of 29
This product manual is subject to change without notice
Copyright © 2008 - 2022 Raditeq B.V.

Page 3 of
Page 3 of 29
This product manual is subject to change without notice
Copyright © 2008 - 2022 Raditeq B.V.
RadiSense® Product Manual
RSS2010E
Table of contents
WARNINGS & PRECAUTIONS 4
The RadiSense® Series 7
RadiSense® Essential System Deliverables: 8
Fibre Optic Cables 9
RadiSupply® plug-In card, rear Panel 9
Laser - safety measures 10
Fibre handling and maintenance 12
Fibre conditions 13
Probe Stand 14
Software conguration 16
Conguring the RadiSense® in RadiMation® 16
How to congure the RadiSense® in RadiMation® 16
How to connect the RadiSense® to the RadiCentre® 17
How to setup the communication of the RadiSense® to the RadiCentre® 18
How to check whether the RadiSense® is properly connected to RadiMation®
19
RadiSense® 10 User correction calibration data 21
Create and upload user correction data with RadiMation® 22
Program the correction le into the probe 24
Enabling the user correction le 24
RadiSense® specications 25
RadiSense® general specications 25
Warranty Conditions 27

Page 4 of
RadiSense® Product Manual
RSS2010E
Page 4 of 29
This product manual is subject to change without notice
Copyright © 2008 - 2022 Raditeq B.V.
WARNINGS & PRECAUTIONS
For cleaning, use a clean, dry cloth (or a damp cloth where needed) and carefully wipe the surface of equipment.
This product does not contain any hazardous substances and complies with the requirements as described in the RoHS
Directive (2011/65/EU)
Read the contents of this product manual carefully and become familiar with the safety markings, the product instructions
and the handling of the system. Please refer to the applicable product manual(s) for further information regarding the
operation and control of the product(s).
This equipment is designed to be used as a plug-in card for the RadiCentre® series modular test systems. Do not use this card on its
own or in combination with any other mainframe. Using this product with any other mainframe can cause harm and will void
warranty.
Only Raditeq qualied maintenance personnel is allowed to perform maintenance and/or repair service on the equipment.
This product contains materials that can be recycled and reused to minimize material waste. At the ‘end-of-life’, specialized
companies can dismantle the discarded system to collect the reusable and recyclable materials. If your product is at its ‘end-
of-life’, please return it to your local reseller or to Raditeq for recycling.

Page 5 of 29
This product manual is subject to change without notice
Copyright © 2008 - 2022 Raditeq B.V.
RadiSense® Product Manual
RSS2010E
The RadiSense® system is a closed loop bre system and therefore classied as a Class 1M laser system according to EN60825-1 and
EN60825-2. In order to provide laser safety in case of a bre failure or accidental disconnection of the bres, the RadiSense® system is
provided with an Automatic Laser Shut down (ALS) as described in the EN60825-2
Never look into the bre optic connectors or bers of the RadiSense probe and/or plugin-card.
As an additional safety measure, the CTR1004B and CTR10007B require an additional laser code to start the system.
This code can be entered using the touchscreen of these RadiCentre® systems.
The RadiCentre Slim requires a laser start procedure where the user has to hold the start button for at least 5 seconds until
ve warning beeps are heard, before the laser is activated.
Make sure that the bre optic cables are installed correctly before activating the system. Do not activate the system if the bre
optic cables show any sign of damage or tampering.

Page 6 of
RadiSense® Product Manual
RSS2010E
Page 6 of 29
This product manual is subject to change without notice
Copyright © 2008 - 2022 Raditeq B.V.
Introduction
Product Introduction
During Radiated immunity testing, the generated eld is measured with a eld probe. Therefore, the overall
accuracy of your radiated immunity test is completely dependent on the accuracy of this sensor. This makes
it the most important instrument to verify the correct eld-strength is achieved during radiated immunity
testing.
The RadiSense® 10 E-eld sensor is the most accurate wideband E-eld sensor available on the market.
Due to it’s small dimensions, the distortion of the eld is minimized. Additionally our laser powered probes
allow uninterrupted measurements without the hassle of empty batteries.
RSS2010E applications:
• Radiated immunity eld monitoring
• Anechoic chambers calibration
• Field homogeneity measurements
• RF radiation-hazard monitoring
• Long term eld monitoring
RadiMation® Software
RadiMation® is the brand independent EMC automation software from Raditeq.
RadiMation® places the EUT in the center The software is designed to perform all
EMC tests required to certify an EUT. This instead of focusing on separate EMC tests.
As a result, our advanced report generation tool makes generating a test report at
the end of all EMC tests as easy as pressing a single button.
Device drivers for all main manufactures are freely made available or are made for
free when required. This allows you choose the beset EMC test equipment for each
specic application, instead of being locked in by a vendor.
RadiMation

Page 7 of 29
This product manual is subject to change without notice
Copyright © 2008 - 2022 Raditeq B.V.
RadiSense® Product Manual
RSS2010E
The RadiSense® Series
Product Characterstics
Laser Powered - The RadiSense 10 is powered by laser light, eliminating the need for batteries. This allows
24-7 continuous testing at maximum performance without the need to change or recharge batteries. In most
instances changing batteries also requires re-calibration.
Superb measurement accuracy – To perform accurate eld measurements, the E-eld sensor dimensions must
be as small as possible compared to the wavelength of the measured signal. There are two reasons for this:
• Firstly,becauselargesensorsizescausethebodysensortoresonateatlowerfrequencies,causingmeasurement
inaccuracies.
• Secondly, to maintain eld uniformity in an anechoic chamber. The smaller the sensor, the better the obtainable
resolution for eld homogeneity measurements.
InTEM andG-TEM cells, sensordimensions above100mm are largecomparedto the dimensionsof thehomogenate
eld area, making them useless in these situations.
The sensor dimensions of the RadiSense® 10 are small relative to other probes, increasing measurement accuracy.
In addition, the RadiSense® 10 is equipped with two (2) antenna elements per axis (six (6) in total) to optimize
isotropic performance and achieve the best overall accuracy.
As part of the ISO9001 quality procedure each RadiSense® 10 eld sensor is factory adjusted and veried. By
default linearity adjustment data is stored inside the RadiSense® 10 eld sensor.
User correction - calibration factors
Additionally, the RadiSense® 10 series has the capabilities to store “user correction factors” inside the eld sensor.
By applying calibration factors in the sensor itself, there is no risk one forgets to apply the frequency correction
data when performing manual or automated tests.
The frequency response correction factors which are determined during an ISO/IEC 17025 accredited calibration
of the probe can be stored inside the eld sensor using the RadiMation® device driver.
Please refer to paragraph 5 of this manual for further instructions.
When the user correction data is stored inside the probe the RadiSense® 10 eld sensor will return corrected
E-eld readings without the need to apply additional corrections. In this case corrections will be applied inside
the probe for each individual axis.

RadiSense® Product Manual
RSS2010E
Page 8 of 29
This product manual is subject to change without notice
Copyright © 2008 - 2022 Raditeq B.V.
RadiSense® Essential System Deliverables:
The RadiSense® Essential sensor can ONLY be delivered as a complete set.
RadiSense® E-Field Probe
Model: RSS2010E - 20 MHz to 10 GHz, 15 measurements/second
RadiSupply® plug-in card
Model: LPS2001B
The electric eld sensor plug-in card to be used in the RadiCentre® system.
RadiCentre® Essential Single slot system
Model: CTR1001S
The Smallest RadiCentre® system usable for single plug-in cards
Fibre extension cable
Model: CBL2001-10m
10-meter bre extension cable with FC/ST connectors on both sides. Fibre
extension cables with dierent length and additional coupling sets can be
purchased separately.
Coupling set
Model: CPL2001A
Coupling set consisting of a FC/PC and ST/PC coupling to connect the
extension bre optic cable to the sensor bre optic cable.
Standard included supporting documentation in the form of:
• USB stick containing:
• The (digital) product manual and Quick Start Guide.
• Factory verication certicate (Raditeq)
• Protective Transport Case
• Hardcopy of the Quick Start Guide.
• Optional - The calibration certicate for the eld sensor
(if an ISO/IEC 17025 accredited calibration certication was ordered).
• Lint free cleaning wipes

Page 9 of 29
This product manual is subject to change without notice
Copyright © 2008 - 2022 Raditeq B.V.
RadiSense® Product Manual
RSS2010E
Fibre Optic Cables
Use an extension bre to connect the sensors to the plug-in card mounted in the RadiCentre®. This extension bre
is a robust duplex bre optic cable and uses dissimilar connectors to avoid faulty connections. The bre optic
cable with FC connectors feeds laser light to the eld sensor. The bre optic cable with ST connectors is used for
bi-directional data communication between the eld sensor and the plug-in card.
To ensure the safe and correct operation of the sensor, only use the RadiSense® with the original supplied or
ordered bres optic cables. Do not use bres optic cables other than those supplied by Raditeq, The use of ber
optic cables from another supplier can lead to permanent damage.
Fibre optic extension cables with dierent lengths up to 100 meters are available on request. Note that for the
probe to correctly function the bre optic cable can be extented by a maximum of two coupling sets and a
maximum total length of 100m.
Contact your local reseller for more detailed information.
RadiSupply® plug-In card, rear Panel
The following connections and indicators are found at the back panel of the RadiSupply® plug-in card:
POWER ON:
This (green) LED lights up when the power is on and the interlock is
closed.
LASER APERTURE:
Connect to the E-eld sensor through the bre extension cable equiped
with FC connectors.
LASER ON:
This red LED lights up when the laser is active.
OPTICAL DATA LINK:
Connect to the E-eld sensor through the bre extension cable equiped
with ST connectors.

RadiSense® Product Manual
RSS2010E
Page 10 of 29
This product manual is subject to change without notice
Copyright © 2008 - 2022 Raditeq B.V.
Laser - safety measures
The RadiSense® uses a high-power laser to supply energy
to a remote measuring device. The wavelength of this laser
is approximately 808nm. This infrared laser is invisible to the
human eye.
During normal operation, exposure to laser radiation is not
possible because the RadiSense® uses a bre-coupled closed loop system with Automatic Laser Shutdown (ALS).
Despite the automatic security measures of the RSS2010E, the precautions below must be followed under all
circumstances.
NEVER look into any of the bre optic connectors. The laser emits a beam that is invisible to
the naked eye. Such light may cause permanent eye damage. Avoid eye or skin exposure to
direct or scattered radiation.
• Assign a ‘laser safety ocer’ in your company. The ‘laser safety ocer’ is responsible for reviewing the
safety precautions.
• Check and connect all bre optic cables before activating the system.
• Do not activate the system if the bre optic cables show any sign of damage or tampering.
• The RadiCentre® system is equipped with a remote interlock system. This remote interlock connection should
be connected to an ‘emergency stop switch’.
• A RED ‘LASER ON’ indicator LED is present on both the front of the RadiCentre as well as on the back of the
RadiCentre (near the FO connectors). These LED’s will be ON when the laser is activated. These LED’s serve
as extra indicator to the operator that one or more lasers are switched on. Do not disconnect bber optic
cables as long as these LED’s are ON.