Scarlet Tech WL-410 Manuale utente

WL-410 Series
User Guide
Wireless
Wind Monitoring System

WL-410 Wireless Wind Alarm System consists of a wind speed sensor and a receiver. The
sensor measures wind speed and transmits data to the receiver. The 3.5-inch LCD display
of receiver shows readings on the main page and the history graphs are shown in different
pages. According to the preset values, the receiver can trigger local alarms such as LED
lights and buzzer, or trigger external indication via 4 built-in relays when measured wind
speeds exceeding the thresholds.
A 868 MHz radio frequency module is installed in the receiver. In WL-410XB model, an
additional Bluetooth Low Energy chip is installed to enable smartphone communication
functionality. WL-410XB sensor output power is stronger (+14 dBm) than WL-410, allowing a
longer transmission distance. An optional 4-20 mA current loop module is available to
WL-410XB for long distance cable transmission.
Overview
Instrument at a Glance
Sensor
Receiver
4 programmable alarm relay outputs
433/868/915 MHz
400m(WL-410)
1,300m(WL-410XB)
Bluetooth
40m
4-20 mA output
(WL-410XB 4-20mA)
Relays
Smartphone App
iOS/Android
External
Indicators
Control Room /
Equipment

Power
When the “Lo battery” appears on the display, replace the sensor battery. Take off the
battery cover (on the bottom) by turning it anti clockwise. Pull out the battery holder and
insert new battery (Lithium 3.6 V AA).
Range
Connection between sensor and receiver goes through free 868 MHz band (optionally
908MHz). Transmission range is up to 500 meters (1300 meters at models XB) - at 10 m
sensor mounting height and when there is no obstacle between the sensor and the display
unit. Inside buildings the range is much shorter. Normally the signal can be received
through two to three walls.
Address
The correspondent sensor address has been preset at the factory. Always set the address
correspond to the address of the sensor from which the receiver will receive data. Address
of the sensor is indicated on the sensor label and in the interior of sensor battery compart-
ment.
Sensor unit switches ON automatically when the anemometer cups revolve. It switches
OFF 6 hours after the anemometer cups stop revolving. In ON state, the sensor updates
wind speed data every two seconds.
Sensor & Receiver
Wind cup
(replaceable)
Battery cap
(O-ring sealed)
Sensor body
Wind cup
(replaceable)
Sensor body
Sensor
Wind vane

Port 1 Port 2 Port 3
Buzzer
Yellow LED for Pre Alarm 2
Up/Mute button
Down/History button
Set menu button
Pre Alarm 2 indicator
Pre Alarm 1 indicator
Max Alarm indicator
Red LED for Max AlarmLCD display
Antenna (SMA connector)
Select the position where the signal reception is strong enough under all conditions.
Connect the enclosed antenna to the SMA connector on the display unit only when the
power supply is switched OFF.
Signal Reception
The reception symbol blinks when the display unit receives a signal from the sensor (every
2 seconds if the reception is good). Signal strength is numerically shown in dB:
-105 dB is approximately the limit where display unit stops receiving
-100 dB and lower means very week signal
-95 ... -90 dB is considered as still acceptable
-85 dB and higher is a good signal
When the sensor doesn't send data (in the OFF state) or when the signal is lost for a more
than 30 seconds, the “No data” notification appears on the display, indicating that the
receiver doesn’t receive data from the sensor (sensor OFF or out of range).
Alarm
The threshold wind speed of Pre Alarm 1, Pre Alarm 2 and Max Alarm can be configured via
setting menus. Please see the section “Settings” for more details.
By holding-down the upper key for 2 seconds you can switch the sound alarm ON and OFF.
The sound alarm is active each time the display unit is turned ON. In the settings menu, the
sound alarm can be disabled permanently.
Receiver

Current wind speed
Signal strength
Temperature Pre alarm 2 threshold
Max alarm threshold
Peak wind bar graph
Max alarm level
Average wind direction indicator
Current wind
bar graph
Wind speed unit
Scarlet WL-410 display consists of 3 pages including: wind speed readings (page 1), wind
speed history graphs (page 2), and maximum alarm history graphs (page 3).
By pressing HIST key to navigate to page 2 and 3 for the wind speed and max alarm history
graphs. To switch back to the first page user can long pressing HIST key for 3 seconds or
inactive for 1 minute.
The contents of display are shown above. Note: in both bar graphs, the max alarm value set
by user is always on the 3/4 bar height. One icon not shown in the photo is data packet
receiving indicator and it flashes by the side the signal strength.
8 hours peak wind speed history graph,
each column presents peak wind speed
during the 5 minutes intervals
First Page
Second Page
.
8 hours Maximum alarm history
graph with 5 minutes intervals.
Third Page
Display

.
Wirings
12 to 24 V DC source
The working voltage of receiver is 12 to 24 V DC. Please plug the power adapter or your own
DC power source to the corresponding jack on the circuit board after opening the front
panel.
Relays for External Alarm Devices
For using relays, connect the connection terminals to external devices.
RELAY 1 : When set value exceed the relay 1 will activate, red light will flash and buzzer will
continuously sound.
RELAY 2 : When wind speed exceeds the value set on pre alarm 2, the relay 2 will activate
and yellow light will start blinking and buzzer will sound with interruptions.
RELAY 3 : When wind speed exceeds the value set on pre alarm 1, the relay 3 will activate.
Please see section “Settings” for more details.
RELAY 4 : Set for alerting that the receiver has no data from the sensor or as 2-zone wind
direction alarm.
4-20mA Industrial Outputs (WL410 XB/4-20mA only)
For using 4-20mA outputs, please connect the wind speed and/or wind direction connec-
tion terminals to external devices.
NOTE: Please make sure the polarity is correct otherwise the receiver will be damaged.
WL-410 XB/4-20mA rear panel wiring connection
12-24V RELAY
4
RELAY
3
RELAY
2
RELAY
1
DC
4-20 mA
WS WD
4-20 mA
12-24V
DC
mA
mA
Please switch off the power before unscrew all 4 plastic screws and remove the front panel

Settings Procedure
Settings List
Settings
1. Hold the SET key to enter the settings menu. If password protection is active, enter the
correct password. The group of settings displays on the display.
2. Select the group by using the up and down keys and press the SET key to view the
selected group.
3. Using the up and down keys select the parameter you wish to adjust and press the SET
key to view the selected parameter. The adjustable parameter blinks.
4. With the up and down keys, adjust the parameter value. Press SET to enter the new
value and move to the next parameter.
5. Exiting the settings menu: hold down the SET key to move back one level. The anemom-
eter also returns to the main screen after 2 minutes of inactivity.
Max Alarm
Pre Alarm 2
Pre Alarm 1
No Signal/WD alarm
General
Relay 1, RED ALARM LIGHT
Relay 2, YELOW ALARM LIGHT
Relay 3
Relay 4
select to set Max Alarm parameters
select to set Pre Alarm 2 parameters
select to set Pre Alarm 1 parameters
select to set No Signal / Wind Direction alarm parameters
select to set general parameters
12-24V
RELAY
4
RELAY
3
RELAY
2
RELAY
1
DC
4-20 mA
4-20 mA
12-24V
DC
mA
mA
WS WD
Settings: Max Alarm (Relay 1, RED ALARM LIGHT, continuous sound)
Settings: Pre Alarm 2 (Relay 2, YELOW ALARM LIGHT, interrupted sound)
Settings: Pre Alarm 1 (Relay 3)
Wind Speed:
ON Delay:
OFF Delay:
Relay:
Wind Speed:
ON Delay:
OFF Delay:
Relay:
Factory preset:
72 km/h
0 s
0 min
Normal
Factory preset:
52 km/h
0 s
0 min
Normal
Setting range:
1 - 50 m/s
0 - 600 s
0 - 60 min
Normal/Inverted
Description:
Max Alarm limit
Minimum time for excess wind speed to activate the Max alarm
Alarm switch OFF delay after wind speed drops below preset level
Inverted operation of relay
Setting range:
1 - 50 m/s
0 - 600 s
0 - 60 min
Normal/Inverted
Description:
Pre Alarm 2 limit
Minimum time of exceed Wind speed to activate the Pre Alarm 2
Alarm switch OFF delay after wind speed drop below preset level
Inverted operation of relay
Wind Speed:
ON Delay:
OFF Delay:
Relay:
Factory preset:
42 km/h
0 s
0 min
Normal
Setting range:
1 - 50 m/s
0 - 600 s
0 - 60 min
Normal/Inverted
Description:
Pre Alarm 1 limit
Minimum time of exceed Wind speed to activate the Pre Alarm 1
Alarm switch OFF delay after wind speed drop below preset level
Inverted operation of relay
Settings: No Signal / Wind Direction Alarm (Relay 4)
Alarm:
Relay:
Zone1:
Zone2:
ON Delay:
OFF Delay:
Factory preset:
No Signal
Normal
315°-45°
135°-225°
0 s
0 min
Setting range:
No Signal / Direction
Normal/Inverted
0°-359°
0°-359°
0 - 600 s
0 - 60 min
Description:
Type of Alarm: No signal alarm or wind direction alarm
Inverted operation of relay
Zone of wind direction alarm ON
Zone of wind direction alarm ON
Minimum time of wind direction inside the alarm zone to activate
the alarm
Alarm switch OFF delay after wind direction leaves the alarm zone

Reset of Display Unit
With upper button pressed at power ON, RESET is performed (all settings except »Sensor
Address«, »WS Cal. Factor« and »WD
Cal.« go to default values).
General Settings
Sensor address
Averaging period
Wind speed unit
Temperature unit
20 mA output
Password
Sound alarm
Language
Wind speed
calibration factor
Bluetooth
Factory Default
Enclosed sensor
2 sec
km/h
° C
180 km/h
No
On
English
+0.0%
On
Settable Range
1…255
2/10/30 sec
m/s, km/h, mph,knots
° C or ° F
10…50 m/s
No
Yes:0000…9999
On/Off
English/French
On/Off
Description
Set the sensor address of your Wind speed sensor
Averaging period for showed wind speed
Unit of displaying wind speed
Unit of displaying Temperature
Wind speed at 20 mA output
4mA = 0 km/h
20 mA = seted value wind speed
* applicable only at model with additional 4-20 mA output
Activation of password protection and setting of password
Switching sound alarm ON and OFF
Language selection
Calibration factor for wind speed(-15,0 ...+15,0% in 0,5% steps)
Power ON/OFF Bluetooth transmitter
* applicable only at WSM W410XB model
Troubleshooting
Symptom
The receiver/display unit cannot read
the sensor
(“No data” appears on the display)
Interrupting and weak sensor signal
Action
- check if the correct sensor address is set on the display unit
- make sure the sensor is not in sleep mode (turn the cups to wake up the sensor)
- check the sensor battery – replace the battery if needed
- check the operation at a reduced distance to the display/receiver
- check for obstructions between sensor and display/receiver unit
- place the sensor or display/receiver in a different position with better signal reception
- reduce distance to the display/receiver
- change or add antenna for the display/receiver unit

Bluetooth Smart Phone App
(WL-410XB and WL-410XB 4-20mA model)
Scarlet provides free iOS and Android Apps. Users can download directly via QR link below:
Before pairing, please turn on the Bluetooth
function on
your mobile device. Make sure the mobile device
with
Android 4.3 or newer with Bluetooth Low Energy
(BLE).
For iPhone, the
device should be at least iPhone 4s or newer.
In Application setting menu always set the address
corresponding to the address of the sensor from
which you
wish to receive data. Address of the sensor is
indicated on
the label and in the interior of the battery compart-
ment.
For example, enter the sensor address showed on
the
label, e.g. 78. Once it’s paired, the data will be
automatically
transmitted to the mobile device.
Smart Phone Applications
Sensor Pairing
iOS Android

Technical Specification
Packing list
• Wind Sensor x1
• Receiver Unit x1
• Power Adapter (input:100-240V, 50/60Hz, 1.5A; output:25V, 2.5A) x1 • Antenna x1
• 3.6V AA Lithium Battery x1
• Display Holder x1
• Waterproof Carrying Case x1
Optional Accessories
• Magnetic Sensor Mounting Bracket
• Wind Sensor Range Extender
• External Antenna
Standard
EMC This instrument was designed in accordance with EMC Standards in force and its
compatibility has been tested in accordance with EN61326-2 (2006).
Spare Parts
• Wind Cups
• Battery(3.6V AA)
Subject to technical modification without notice
Sensor & receiver
Measurement range
Unit
Resolution
Accuracy
Averaging period
Operating voltage
Power consumption
Transmission distance
Distance with YAGI antenna
Relay contact rating
Temperature operating range
Sensor battery
Sensor battery life
Audible alarm
Antenna input (receiver)
Output
Relay
Output
Mechanical
Bearings
Sensor housing
Sensor cups (replaceable)
Receiver housing
Sensor dimension
Receiver dimension
Mounting
Bluetooth LE
2 x precision stainless steel ball
AL/PVC
PA (Nylon)
ABS, IP65
Height 210 mm, cup-to-cup diameter 120 mm
150 x 80 x 55 mm
Mounts on a pipe with ø20 mm outside diameter
WL-410 WL-410XB WL-410XB/4-20mA
-
-
-
Yes. Range is up to 40 m.
1300 m
6.5…10.0 km
3 years
0.6…50.0 m/s
m/s, km/h, knots, mph
0.1 m/s
+/-3%
Selectable 2 s, 10 s, 30 s
12…24 V DC
300 mA (max.)
1 A/12 V DC; 1 A/24 V DC
-25…60° C
3.6 V AA Lithium batteries
85…90 dB
50 ohm, SMA connector
4 x relays, 2 A / 24 V
4…20 mA
4 mA=0 m/s
20 mA=10…50 m/s
(programmable)
400 m
1.5…2.4 km
5 years
Altri manuali per WL-410
1
Questo manuale è adatto per i seguenti modelli
3
Indice
Altri manuali Scarlet Tech Strumento di misura

Scarlet Tech
Scarlet Tech ST-109 Manuale utente

Scarlet Tech
Scarlet Tech ST-25D Manuale utente

Scarlet Tech
Scarlet Tech ST-130 Manuale utente

Scarlet Tech
Scarlet Tech WR-3 Manuale utente

Scarlet Tech
Scarlet Tech WindPro Manuale utente

Scarlet Tech
Scarlet Tech ST-11D Manuale utente

Scarlet Tech
Scarlet Tech TWL-1S Manuale utente

Scarlet Tech
Scarlet Tech ST-109 Manuale utente

Scarlet Tech
Scarlet Tech WR-3 Plus Manuale utente