13
MENU continued from page 12
Pressing the MENU (B) button will display the date and time of
that session. Pressing the MENU (B) button again will display
thegas reading.Pressing theSAVE/ZERO(C)button willallow
youto scrollthrough allother gasreadings. Pressthe POWER/
MUTE (A) button to return to SESSION (No.) and pressing the
SAVE/ZERO (C) button will allow you to review all previously
saved SESSION’s by date. Press the MENU (B) button to re-
view the gas data. Pressing the POWER/MUTE (A) button 3
timeswillreturn youtothe USERMENU.Pressingthe POWER/
MUTE(A) buttononce morereturns youto theworking display.
The number of stored session log saves is factory set at 6. It
canstore upto 16 bychanging factorysettings (contactfactory
forinstructions).
BUMPTEST
From the working display access the menu by pressing and
holding the MENU (B) button until the top line of the display
reads USER MENU. Press and release the SAVE/ZERO (C)
button until the bottom line displays BUMP TEST. Prepare
100ppmCO forapplicationto theinstrument.Attachthe hoseto
theinlet connection. Turnon thegas. Press theMENU (B)but-
ton. EACH GAS must read 80% of calibrated value within 45
seconds. The readings are on the left and the timer is on the
right side of the display. If the instrument passes, the display
will read BUMP TEST PASSES and a beep will sound. If the
instrumentfails,the display willread BUMPTESTFAILEDand
abeepwillsound.AttheendofanybumptestpressthePOWER/
MUTE (A) button to return to the working display.
(Continued on page 14) 14
SMART-CAL
From the working display access the menu by pressing and
holding the POWER/MUTE (A) button until the top line of the
displayreadsUSERMENU. PressandreleasetheSAVE/ZERO
(C) button until the bottom line displays SMART CAL. Place
the instrument into the cradle provided on the left side of the
Smart-CalCalibration Station.Attachthetubing fromthestation
to the inlet side of the instrument. Press the center button. The
displaywill show“SMART CAL Communicating”and thepump
will turn off. Select the test from the Smart-Call Station to be
performed. At the end of the test the instrument will beep 3
times and display PASS or FAIL. Retry the test if necessary by
pressingtheproperbuttonon theSmart-CalStationagain.Press
the left button to return the working display, remove the tubing
andreturninstrumentto serviceorsendinstrument totheproper
placefor repairper company procedures.
SHORTCUT TO ACCESS SMART-CAL: Place theinstrument
intothe cradleprovidedon theleftsideofthe Smart-Calcalibra-
tionStation.Attach thetubingfrom thestationto theinlet sideof
the instrument. While in the working display press the power
button for 2-3 seconds and release. The display will show
“SMART CAL Communicating” and the pump will turn off. Per-
form all tests as described in the SMART CAL section.
MENU continued from page 13