Chapter 2 Panel Overview
Note:
1. Accuracy of D and Q are delta absolute deviations, the rest are delta percent deviations;
2. Suffix withX is the actual measured value of this parameter, suffix with max is the maximum
value, suffix with min is the minimum value;
Range Hold
Parameter Range
Auto Range0 Range1 Range2 Range3 Range4
Cmax 80µF/f 1000pF/f 0.1µF/f 1µF/f 10µF/f 80µF/f
Cmin 150pF/f 150pF/f 1900pF/f 10nF/f 100nF/f 1µF/f
Lmax 159H/f 159H/f 25.3H/f 2.53H/f 253mH/f 25.3mH/f
Lmin 0.32mH/f 2.53H/f 0.25H/f 25.3mH/f 2.53mH/f 0.32mH/f
Zmax 1MΩ1MΩ159kΩ15.9 kΩ1.59kΩ159Ω
Zmin 1.59Ω15.9kΩ1.59kΩ159Ω15.9Ω1.59Ω
Z
max = Rmax; Zmin = Rmin
3. In order to maintain good test accuracy, proper open and short circuit corrections should
be taken under current measurement condition and measurement fixture.
4 Environment
1.Please do not operate the instrument under the following environment conditions, as
any of them will directly affect measuring precision or damage the instrument:
(1) Please do not operate the instrument in the places where is vibrative, dusty,
under direct sunlight, or where there is corrosive air.
(2) Although the equipment has been specially designed for reducing the noise
caused by AC power, a place with low noise is still recommended. If this cannot
be arranged, please make sure to use power filter for the instrument.
2.Please store the instrument in the place where temperature is between -25°C and
50°C. If the instrument will not be put in use for a while, please have it properly packed
for storing.
3.Operation environment for securing measuring precision requirement:
Temperature: 0°C ~ 40°C
Humidity:≤85%R.H.
4.Pre-heat the instrument about 10 minutes before starting measuring.
6