Instructions
1. Install two AAA batteries into the compartment on the back of the meter
2. Press and hold power button until the meter turns on. Wait 5 seconds before taking your
first measurement.
3. Check the three segment battery indicator in the top, right screen. If the meter’s power is
low, replace the battery. Low power may effect the measurement function.
4. Measuring static:
a. The meter should be 25 mm (.98inch) away from and parallel to the surface to be
measured.
b. The WD button turns on the LED guide.
c. Press the HOLD button to keep the measured data on the display after a test is
taken. Press the HOLD button again to resume detection mode.
Notes:
1. Wait 5 seconds after powering on before taking first measurement
2. Do not zero out the meter in presence of static. This may effect the accuracy.
3. Make sure the static sensor if clean. Do not insert anything into the sensor. This can cause
damage.
4. Do not measure static charges over 2000volts. This can damage the meter. Note: It typically
takes a charge of greater than 2,000 volts for a person to feel the “zap”. If you can sense the
charge, do not measure it.
5. The meter is sensitive to drops and collisions. This can cause malfunctions.
6. Do not disassemble the meter. If there is an problem, contact Transforming Technologies.
Maintenance.
1. The meter should be clean and free of dirt. Clean of meter with a soft cloth. IPA diluted with
water can be used for tougher stains
2. The EFM250 is factory calibrated. It is recommended to verify calibration once per year.
Contact Transforming Technologies for more information.
3. The meter has a One Year Warranty for manufactures defects.
Troubleshooting:
Abnormal value or error during
measurement
Too close to high voltage equipment,
ionizers or other strong
electromagnetic equipment
Remove/turn off equipment that is the
source of the interference
Detection distance setting error
Reposition meter within 25mm of
surface
The meter has an odor Component burning Stop testing. Contact Transforming
Technologies for evaluation
If there are problems that are not listed, contact Transforming Technologies for evaluation