3000 A RANGE
Primary Current 150 A 600 A 3000 A
Accuracy % 1.5 % 0.75 % 0.5 %
Phase shift 1.5 ° 0.75 ° 0.5 °
Load Impedance: ≤0.4 Ω
Overload: 3600 A for 10 min
Ampere-Second Product: 80 A·s
Accuracy: Per IEC 185-26-27
Class 0.5 (5 V·A) from 48 Hz to 1000 Hz
(*Reference conditions: 23 °C ± 3 K, 20 %RH to
85 %RH, 45 Hz to 65 Hz, external magnetic eld
<40 A/m, no DC component, no external current
carrying conductor, test sample centered.)
Frequency Range: 30 Hz to 1500 Hz; cur-
rent derating above 1500 Hz for continuous
use
Load Impedance: ≤2 Ω
Working Voltage: 600 VAC
Common Mode Voltage: 600 VAC
Inuence of Adjacent Conductor:
<0.005 A/AAC
Inuence of Conductor in Jaw Opening:
1.5 % ± 0.2 A on 1000:5 range
1 % ± 0.2 A on 2000:5 range
1 % ± 0.2 A on 3000:5 range
MECHANICAL SPECIFICATIONS
Operating Temperature:
14 °F to 122 °F (-10 °C to 50 °C)
Storage Temperature:
-13 °F to 176 °F (-25 °C to 80 °C)
Inuence of Temperature:
<0.1 % per 10 K
Jaw Opening: 3.54 in (90 mm)
Maximum Conductor Size:
Cable: 2.52 in Ø max (64 mm)
Bus bar: 1.97 in x 5.31 in (50 mm x 135 mm)
2.52 in x 3.94 in (64 mm x 100 mm)
Envelope Protection:
IP 20 (IEC 529)
Drop Test:
1 m (IEC 68-2-27)
Mechanical Shock:
100 g (IEC 68-2-27)
Vibration:
10/55/10 Hz, 0.15 mm (IEC 68-2-6)
Polycarbonate Material:
Handles: 10 % berglass charged
polycarbonate UL 94 V0
Jaws: ABS UL 94 V2
Dimensions:
4.72 in x 12.40 in x 1.89 in
(120 mm x 315 mm x 48 mm)
Weight:
2.65 lb (1200 g)
Colors:
Dark gray handles with red jaws
Output:
Two standard safety banana jacks (4 mm)
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation
between primary or secondary and outer
case of handle upon IEC 1010-2-32
Common Mode Voltage:
600 V CAT III, Pollution: 2
ORDERING INFORMATION
Current Probe JM865A........ Cat. #2110.88
Accessories:
Adapter - Banana (Female) to BNC (Male)
(XM-BB) (600 V CAT III) ........ Cat. #2118.46
Lead - set of 2, 5 ft (red/black) PVC
with 4 mm ST/ST plugs .........Cat. #5100.07
OPERATION
Please ensure that you have read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model JM865A
■Connect the black and red terminals to the Ampere AC range of your DMM or current
measuring instrument. Select the appropriate current range. If the current magnitude
is unknown, select the highest range (3000 AAC / 5 AAC) on the switch probe inside
the handle. Clamp the probe around the conductor with the arrow pointed toward the
load. Select the lower range until you obtain the best resolution. Read the value display
on the DMM and multiply it by the range selected. If the reading equals 2.59 A on the
1000/5 range, the current owing through the probe is 518 AAC (2.59 A x 200).
■For the best accuracy, carefully center the conductor inside the probe jaw, avoid if
possible, the proximity of other conductors which may create noise.