AEMC 6240 Manuale utente

MICRO-OHMMETER 6240
ENGLISH User Manual
NOTE: For instruments with serial numbers
lower than 116339SEDV, terminals C1 and P1
are red, and C2 and P2 are black


Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
certifies that this instrument has been calibrated
using standards and instruments traceable to
international standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
An NIST traceable certificate may be
requested at the time of purchase, or obtained
by returning the instrument to our repair and
calibration facility, for a nominal charge.
The recommended calibration interval for this
instrument is 12 months and begins on the date of
receipt by the customer. For recalibration, please
use our calibration services. Refer to our repair
and calibration section at www.aemc.com.
Serial #: ________________________________
Catalog #: 2129.80
Model #: 6240
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.
d.b.a AEMC®Instruments
www.aemc.com

Table of Contents
1. INTRODUCTION ................................................................................. 4
1.1 International Electrical Symbols................................................5
1.2 DenitionofMeasurementCategories .....................................5
1.3 ReceivingYourShipment..........................................................5
1.4 OrderingInformation.................................................................6
1.4.1 AccessoriesandReplacementParts............................6
2. PRODUCT FEATURES......................................................................... 7
2.1 Description................................................................................7
2.2 Applications ..............................................................................8
2.3 KeyFeatures ............................................................................8
2.4 ControlFeatures .......................................................................9
2.5 ButtonFunctions.....................................................................10
2.6 DisplaySymbols ..................................................................... 11
3. SPECIFICATIONS............................................................................. 12
3.1 Electrical .................................................................................12
3.2 Mechanical..............................................................................13
3.3 Environmental.........................................................................14
3.4 Safety......................................................................................14
4. OPERATION..................................................................................... 15
4.1 PrecautionsBeforeUse..........................................................15
4.2 InstrumentConguration(SET-UPmode) ..............................15
4.3 ResistanceMeasurements .....................................................16
4.3.1 MeasurementReadings..............................................17
4.3.2 ConnectionsandReadings.........................................17
4.3.3 Test Lead Connection .................................................18
4.4 RepetitiveMeasurements .......................................................18
4.5 MeasuringVeryLowResistance ............................................19
4.6 AutomaticRecording ..............................................................20
4.7 StoringResultsintoMemory...................................................21

Micro-Ohmmeter Model 6240
3
4.8 RecallingResultsfromMemory..............................................22
4.9 ErasingMeasurementsfromMemory.....................................23
4.9.1 ErasingSelectedMeasurements................................23
4.9.2 ErasingAllMeasurements ..........................................23
5. DATAVIEW®SOFTWARE................................................................... 24
5.1 InstallingDataView®................................................................24
5.2 Micro-ohmmeterControlPanel...............................................26
6. TROUBLESHOOTING ........................................................................ 27
6.1 ErrorMessages ......................................................................27
6.1.1 VoltagePresent ..........................................................27
6.1.2 Overrange...................................................................28
6.1.3 NoisyMeasurement....................................................28
6.1.4 Overheating ................................................................28
7. MAINTENANCE................................................................................ 29
7.1 Warning..................................................................................29
7.2 Cleaning..................................................................................29
7.3 Charging/RechargingtheBattery ...........................................30
7.4 BatteryandFuseReplacement ..............................................31
RepairandCalibration ....................................................................32
TechnicalandSalesAssistance......................................................32
Limited Warranty .............................................................................33
WarrantyRepairs ............................................................................33

4
Micro-Ohmmeter Model 6240
CHAPTER 1
INTRODUCTION
WARNING
These safety warnings are provided to ensure the safety of
personnelandproperoperationoftheinstrument.
• Thisinstrumentisprotectedfromaccidentalvoltagesofnot
more than 50V with respect to earth in measurement Cat-
egoryIII.
• Donotattempttoperformanytestswiththisinstrumentuntil
youhavereadtheusermanual.
• Tests are to be carried out on de-energized circuits only!
Never connect the unit to a live circuit.
• Make sure that none of the input terminals are connected
andthattheswitchissettoOFFbeforepluggingintheAC
powertorechargethebatteryoftheinstrument.
• Makesuretheinternalbatteryisfullychargedpriortotesting.
If the instrument has been left unused for several months,
rechargethebattery.
• If the case needs cleaning, do not use any alcohol or oil
based cleaners. Preferably use soapy water with a damp
clothorsponge.Donotimmersethe6240micro-ohmmeter
inwater.
• Use connection accessories of which the overvoltage cat-
egory and the service voltage are greater than or equal to
those of the measuring instrument (50V CAT III). Use only
accessoriesthatcomplywithsafetystandards(IEC61010-2-
031).
• The test leads and measuring wires must be in good
conditionandshouldbereplacedifthereisanyevidenceof
deterioration(insulationsplit,burnt,etc.).
• Troubleshooting and metrological verication procedures
mustonlybeperformedbyqualied,approvedpersonnel,or
thefactory.

Micro-Ohmmeter Model 6240
5
1.1 International Electrical Symbols
Signifies that the instrument is protected by double or reinforced insulation.
This symbol on the instrument indicates a WARNING and that the operator must refer
to the user manual for instructions before operating the instrument. In this manual,
the symbol preceding instructions indicates that if the instructions are not followed,
bodily injury, installation/sample and/or product damage may result.
Risk of electric shock. The voltage at the parts marked with this symbol may
be dangerous.
When disposed of, this product must be sorted for the recycling of electrical
and electronic equipment in accordance with WEEE directive 2002/96/EC.
1.2 Definition of Measurement Categories
CAT IV: For measurements performed at the primary electrical supply
(<1000V) such as on primary overcurrent protection devices,
ripplecontrolunits,ormeters.
CAT III: For measurements performed in the building installation at
the distribution level such as on hardwired equipment in xed
installationandcircuitbreakers.
CAT II: For measurements performed on circuits directly connected to
the electrical distribution system. Examples are measurements
onhouseholdappliancesorportabletools.
1.3 Receiving Your Shipment
Uponreceivingyourshipment,makesurethatthecontentsareconsistent
withthepackinglist.Notifyyourdistributorofanymissingitems.Iftheequip-
mentappearstobedamaged,leaclaimimmediatelywiththecarrierand
notifyyourdistributoratonce,givingadetaileddescriptionofanydamage.
Savethedamagedpackingcontainertosubstantiateyourclaim.
NOTE: Charge the instrument fully before use.

6
Micro-Ohmmeter Model 6240
1.4 Ordering Information
Micro-ohmmeter Model 6240............................................ Cat. #2129.80
Includes extra large tool bag, set of two 10ft Kelvin clips (10A - Hippo), set of two 10ft Kelvin
probes (1A - Spring Loaded), optical USB cable, US 115V power cord, two spare fuses
(12.5A), NiMH rechargeable battery pack, and USB stick supplied with product user manual
and DataView®software.
1.4.1 Accessories and Replacement Parts
Kelvinclips(10A-Hippo),10ftcolor-codedleads
withspadelugterminations................................................. Cat. #1017.84
Kelvinclips(10A-Hippo),20ftcolor-codedleads
withspadelugterminations................................................. Cat. #2118.70
KelvinProbes(1A-SpringLoaded),10ftwith4mmbanana
plugtermination(includessetof5forkterminals)............... Cat. #2118.73
KelvinProbes(1A-SpringLoaded),20ftwith4mmbanana
plugtermination(includessetof5forkterminals)............... Cat. #2118.74
KelvinProbesPistolGrip10ft(10A)SpringLoaded........... Cat. #2118.75
KelvinProbesPistolGrip20ft(10A)SpringLoaded........... Cat. #2118.76
KelvinProbes10ft(10A)SpringLoaded ........................... Cat. #2118.77
KelvinProbes20ft(10A)SpringLoaded ............................ Cat. #2118.78
KelvinClips10ft(1-10A)..................................................... Cat. #2118.79
KelvinClips20ft(1-10A)..................................................... Cat. #2118.80
Thermo-HygrometerModelCA846 .....................................Cat. #2121.24
OpticalUSBCable...............................................................Cat. #2135.41
KelvinClips25ft(200A-Hippo)..........................................Cat. #2129.72
KelvinClips50ft(200A-Hippo)..........................................Cat. #2129.73
Battery–ReplacementNiMH6V9000mAH
forModels6240&6250 ......................................................Cat. #2129.91
Fuse,setof5,12.5A/500V,6.3x32mm ............................. Cat. #2129.92
ExtraLargeClassicToolBag ..............................................Cat. #2133.73
Inverter–12VDCto120VAC200Wattforvehicleuse......... Cat. #2135.43
115VPowerCordUS .......................................................... Cat. #5000.14
Order Accessories and Replacement Parts Directly Online
Check our Storefront at www.aemc.com/store for availability

Micro-Ohmmeter Model 6240
7
CHAPTER 2
PRODUCT FEATURES
2.1 Description
The Micro-ohmmeter Model 6240 is used to perform low resistance
measurementsfrom5μΩto 400Ω.Therearesixrangeswithtestcurrents
from10mAto10A.
Thefrontendofthemicro-ohmmeteremploysafour-wireKelvincongu-
ration,whicheliminatestestleadresistanceforameasurementaccuracy
of0.25%.Abuilt-incircuitltersoutACsignals.
TheMicro-ohmmeterModel6240ispackagedinasealedeldcasewell
suitedforshopandelduse.Powerissuppliedbyalong-lifeNiMHbattery
packwithabuilt-inrecharger(110/220V).
Thelarge,easy-to-readliquidcrystaldisplayis2.25x4.00".Itdisplaysthe
valueofresistance,currentorvoltagetest,polarityandbatterycharge.
Foroperatorsafetyandinstrumentprotection,themicro-ohmmeterisfuse
protectedattheinputs.Onefuse,accessiblefromthefrontpanel,protects
againststoredenergyininductiveloads.
Enhanced internal circuitry protects against possible inductive kickback
whenthecurrentisshuto.
Abuilt-inthermalswitchprotectsthemicro-ohmmeteragainstoverheating
onthe10Arangewhenincontinuoususe.

8
Micro-Ohmmeter Model 6240
2.2 Applications
Someofthemorepopularandmostfrequentusesofthemicro-ohmmeter
areinapplicationsfor:
• Checkingmetalliccoatingresistance,especiallyinaeronautics
• Groundconnectionsandcontinuitymeasurement
• Resistancemeasurementsonmotorsandsmalltransformers
• Contactresistancemeasurementsonbreakersandswitches
• Componentmeasurement
• Electricalcableresistancemeasurement
• Mechanicalbondtests
• Wire to terminal connections
• Aircraftandrailbonds
2.3 Key Features
• Measuresfrom5µΩto 400Ω
• Testcurrentselectionfrom10mAto10A
• Manualtemperaturecompensation(withDataView®software)
• Frontpanelpolarityreversefunction
• Storesupto99testresults
• Operatorsafetybyautomaticdischargeofresidualchargeonthe
equipmentundertest
• Instantaneous,continuousormultipletestoperation
• Autostoreofmultipletestresults
• Internal,rechargeablebatteriesconductupto850,10Atests
• Abuilt-inbatterypackrechargerrechargesthebatteriesby
connectingtotheACline(110V-230V,50/60Hz)usingastandard
line cord
• 4-Wiremeasurementwithautomaticcompensationof
undesirablevoltagesandleadresistance
• Largemulti-functionbacklitdisplay
• Directdisplayofthemeasurementwithunits,rangeandtest
current
• Rugged,sealedcase
Altri manuali per 6240
4
Indice
Altri manuali AEMC Strumento di misura