6. Working Principle
The medium flows through a specially formed flow-chamber and causes thereby
an impeller to rotate. This rotating motion is scanned opto-electronically in a
contactless manner and converted into an asymmetrical frequency signal.
Optionally, a frequency divider with symmetrical output is available. The
generated frequency is proportional to the flow rate. The impeller contains
sapphire-bearings, due to which, high linearity and long life of device are
ensured.
7. Mechanical Connection
·Make sure that the actual flow rate corresponds with the measuring range of
the device. The measuring range (full scale value) can be read off from the
type label.
Note! During measuring range over-stepping (more than 20%), damage to
the axle bearings can occur, which impairs the function of the flow meter.
·Make sure that the permitted max. operating pressure and temperature of the
device DPL are not violated.
·Remove all the transportation-safety locks etc. and ensure, that there exist no
packing-material parts inside the unit.
·These units can be mounted without regard to position. Thereby, the flow must
always take place in the direction of the arrow.
·In particular, attention must be paid to flow from top to bottom, such that the
flow chamber is completely filled with the liquid. Additional inlet and outlet
lengths are not required.
·The sealing of connection screw joints is succeeded through Teflon tape or
similar material.
·During installation of the unit, attention must be paid to see that no high
pressure or tensile loads exert on the connection screw joints. We recommend
to fasten an inlet and outlet pipe mechanically at a distance of approx. 50 mm
from the connection screw joint.
·If possible, after mechanical installation, check whether the sealing of
connection-joints/piping is adequate.