Microtech 110184008 Manuale utente

MICROTECH
WIDE RANGE
COMPUTERIZED MICROMETER
Calibration ISO 17025:2017 ISO 9001:2015
USER MANUAL

MICROTECH
Clean the oil from the measuring surfaces of micrometer by cloth.
Turn on micrometer by pressing SWITCH ON button.
Check the setting display to zero. Contact with each other (or with setting gauge) measuring surfaces of the anvil and the
spindle, using the ratchet and take a measurement.
Rotation the micrometer screw in the direction of convergence measuring surfaces using the ratchet.
For zero setting, press the ZERO button on Touchscreen.
SPECIFICATION
Item No Range Resolution Accuracy Micrometer
head range
Protection
class Indication Data output Calibration
mm inch mm inch
μ
mmm Wireless ISO 17025
110184008 300-400 12-15”
0,0001 0.00001”
±5 25 IP-65
computerized
MICS system
with 1,54”
touchscreen
max 50m +
110185008 400-500 15-20” ±5 25 IP-65 max 50m +
110186008 500-600 20-24” ±6 25 IP-65 max 50m +
110187008 600-700 24-28” ±7 25 IP-65 max 50m +
110188008 700-800 28-32” ±8 25 IP-65 max 50m +
110189008 800-900 32-36” ±10 25 IP-65 max 50m +
110181108 900-1000 36-40” ±10 25 IP-65 max 50m +
110181208 1000-1200 40-48” ±14 50 IP-65 max 50m +
110181408 1200-1400 48-56” ±16 50 IP-65 max 50m +
110181608 1400-1600 56-64” ±20 50 IP-65 max 50m +
110181808 1600-1800 64-72” ±24 50 IP-65 max 50m +
110182108 1800-2000 72-80” ±30 50 IP-65 max 50m +

MICROTECH
For charging connect USB cable
Charging process indicating
on Battery status
(switch on device)
BUILT-IN BATTERY - rechargeable Li-Pol battery
MAIN INFO
CHARGING
PRECISION MEASUREMENTS
MICROTECH computerized micrometer with 0.1
resolution can be used for high precisoт measurements
in manufacturing, automotive medical industry and for
calibration
COLOR GO/NOGO INDICATION
Easy and quick limits control.
It’s possible setup max and min value throw menu.
Color indication reduce possibility of error during
control
INTERNAL MEMORY
Computerized micrometer has memory manger with
possibility saving data to folders. Separate Data’s or Folders
can be transfer Wireless to MICROTECH Software (Windows,
Android,iOS)
FORMULA MODE
In Formula mode we use Ax2+Bx+C
Main using of this function is:
- Preset function
- Magnification of measured value
For customers request Formula can be updated
ERROR COMPENSATION
It’s possible to make mathematical error
compensation. Just re-calibrate micrometer with
gauge block set and save results. Micrometer will
indicate measuring results include re-calibration.
TEMPERATURE COMPENSATION
Temperature compensation function gives
possibility recalculate details sizes with high or low
temperatures to normal (20C conditions). Available
recalculation for 7 materials
Parameters
LED display color 1,54 inch
Resolution 240x240
Indication system MICS 3.0
Power supply Rechargeable Li-Pol battery
Battery capacity 450-800 mAh (depends of device)
Charging port micro-USB
Case material Aluminium
Buttons Switch (Multifuntional), Reset
Wireless data transfer up to 50m on open area
TECHNICAL DATA

MICROTECH
Saved values
Basic STATISTIC functions:
MAX - maximum saved value
MIN - minimum saved value
AVG - average value
D - variance VAR(x)
Each value can be
Deleted or Send to
application Button ALL select all values
All values can be Deleted
or Send to application
MEMORY settings
Selecting Standard mode or
No saving data to internal memory
Memory sub-menu
MEMORY
WIRELESS DATA TRANSFER
Switch on Wireless data transmition
Activating multifunctional button to save value to
memory or send data by touchscreeb ur button push
Activate WIRELESS data transfer in WIRELESS menu
Connect instrument with MICS indication system to Tablet or PC
Send data tp Tablet or PC:
- by Touchscreen
- by Multifactinal button psuh (activted in Wireless menu)
- by Timer (activated in Timer menu)
- from Internal Memory
For save measuring data to internal caliper’s memory touch data area
on screen or short button push. You can view saved data throw menu or
send Wireless connection to Windows PC, Android or iOS devices.

MICROTECH
DOWNLOAD
DOWNLOAD
TRANSFER DATA TO
EXCEL OR OTHER
TABLE EDITORS
TRANSFER DATA TO
CAD, SPC OR OTHER
SOFTWARE
IMPORT FROM FILE OR CAMERA
SAVING
GRAPH
QUALITY
CONTROL
RESULTS
VOICE
MODE
SAVING DATA
TO XLS. CSV
FORMATS
WIRELESS CONNECTION
USB-dONgLE
SAVING DATA
TO XLS. CSV
FORMATS
SAVING DATA
TO XLS. CSV
FORMATS
SAVING
GRAPH
VOICE
MODE
iOS
up to 50m
DATA TRANSFER DISTANCE FOR INTELLIGENT INSTRUMENTS
•All MICROTECH Intelligent instruments has internal memory.
You can collect data to memory and when it will be neccesary
transfer all data to the peripheral device with no data loosing.
•It’s possible transfer data, like in Wireless instruments up to 15 m
or up 50 m with external antenna
•USB port can be used to transfer data directly to Windows PC
DOWNLOAD
CONNECTIONWIRELESS DATA TRANSFER

MICROTECH MENU NAVIGATION
LIMITS
color Go/NoGo
MAX/MIN
indicating max or min value
FORMULA MODE
Ax2+Bx+C
ERROR COMPENSATION
mathematical correction
TEMPERATURE COMPENSATION
manual correction
RESOLUTION
selected up to 0.0001mm
WIRELESS
modes of data transmition
USB
USB connection configuration
DEVICE SETTING
pin, reset
SOFTWARE
link to download app
MEMORY SETTING
CALIBRATION DATE
Battery level
Actual value
Memory
Zero setting
Colour can be changed
according function
MENU button
MENU CONFIGURATION
MAIN MENU
MEMORY WITH STATISTIC
USE TOUCHSCREEN TO OPERATE MENU

MICROTECH
LIMITS GO/NOGO
MAX - NoGo more MAX value
MIN - NoGo less MIN value
RANGE - Go between MIN & MAX
NONE - disactivating mode
MENU SETTING MAIN SCREEN INDICATION
MAX/MIN
MENU SETTING MAIN SCREEN INDICATION
FORMULA
Formula:
Ax2+Bx+C
MENU SETTING EXAMPLE OF FORMULA USING
TIMER
Select timer period
Reset to disactivate mode
MENU SETTING
Standard measuring A=0, B=1, C=0
PRESET 50mm A=0, B=1, C=50
Value magnification x3 A=0, B=3, C=0
Square value A=1, B=0, C=0
NoGo Go
MAX value
MIN value
zero values
MAXMIN
Colour indication limits
Indication and saving MAX or MIN values
ENABLE - activating mode
MAX - indicating MAX measured value
MIN - indicating MIN measured value
Long push digit
to select value
FORMULA Ax2+Bx+C
Reset to zero arguments
Saving data to memory or sending Wireless by Timer
MICS SYSTEM FUNCTIONS

MICROTECH
MATHEMATICAL ERROR COMPENSATION
MENU SETTING
TEMPERATURE COMPENSATION
MENU SETTING
RESOLUTION
MENU SETTING MAIN SCREEN INDICATION
EXTRA (Axis mode)
MENU SETTING
Resolution selecting
Selecting normal or inverted axis mode
(+/- value indication)
Selecting resolution of value on main screen and memory
Reading base reconnection for some
types of calipers
mm/inch conversion
Re-calculation detail size to calibration conditions (20
°
C)
Manual Temperature setting
7 types of material for selection
7 types of material:
- glass, quartz, solid alloys
- granite, graphite
- Fe, NI, Fe alloys
- stainless steel
- cooper, Cu alloys
- alluminium, al alloys
- acrylic sheet
Add new point for
error compensation
Select all points actual values
correct values by up&down button
confirm point correction
Mathematical correction points with big error to normal values
use gauge blocks or measuring machine for points correction
MICS SYSTEM FUNCTIONS

MICROTECH
WIRELESS CONNECTION
MENU SETTING
USB CONNECTION
MENU SETTING
PIN & RESET
MENU SETTING
DISPLAY SETTING
MENU SETTING
- Sleep OFF (15 s low brightness off, sleep off)
- Sleep 15s (15 s low brightness on, sleep off)
- Sleep ON (15 s low brightness on, sleep on)
Switch on Wireless data transmition
Activating multifunctional button to save value to
memory or send data by touchscreeb ur button push
Activating or disactivating wireless data transfer
Display rotation 0
°
, 90
°
, 180
°
, 270
°
Brightness level
Activating PIN to device setting RESET to factory settings
USB connection mode under construction
DISPLAY ROTATION, BRIGHTNESS, SLEEPING MODE
MICS SYSTEM FUNCTIONS
actual values
correct values by up&down button
confirm point correction

MICROTECH
MEMORY SETTINGS
MENU SETTING
LINK TO SOFTWARE
MENU SETTING
CALIBRATION DATE
MENU SETTING
DEVICE INFO
MENU SETTING
Calibration date info
Information about device
- device name
- firmware version
- MAC adress for easy connection
QR Link to MICROTECH web site page
with MDS Software download
- Android, iOS, Windows versions
- Free and Pro versions
- Manuals
Standard - saving values to memory
OFF - disable saving to memory
iOS
Push for change date info
Activating or disactivating saving values to memory
MICS SYSTEM FUNCTIONS
Questo manuale è adatto per i seguenti modelli
10
Indice
Altri manuali Microtech Strumento di misura

Microtech
Microtech 110180027 Manuale utente

Microtech
Microtech e-DIAGNOSTIC Manuale utente

Microtech
Microtech 110180254 Manuale utente

Microtech
Microtech 120126131 Manuale utente

Microtech
Microtech SUB-MICRON 1250511006 Manuale utente

Microtech
Microtech 1106500258 Manuale utente

Microtech
Microtech 120125131 Manuale utente

Microtech
Microtech 14109805 Series Manuale utente

Microtech
Microtech 110181027 Manuale utente