Input
+
*
Display scaling whit minimum input signal. e.g. 4 mA.
Countingdownwardsthedisplay’sminimumvalueisbeingsetfornegative(-).
Look for e.g. page 7. (Zero shift for non scalable inputs)
Display scaling whit maximum input signal. e.g. 20 mA.
Countingdownwardsthemaximumvalueofthedisplayis being set for negative (-) .
Look for e.g. page 7. (multiplier function for mV- and R inputs)
InPut
DEC
Lo
Hi
OPErAt
TimE
StArt
Halt
rESEt
FiLt
dISP
PEAK
CALib
SEtCod
Configurationmenufunctions
Input selection. See chapter 'Selection of input'.
Decimal selection. Select 1..5
FL_nor Rate-of-change (flowrate) Scaled display
Empty (-) 20-4 mA, Fill (+) 4-20 mA e.g. 4..20 mA = 0..100,0%
nor_FL Scaled display without time factor Rate-of-change(Flow/timeunit)
to_nor Totalizer, e.g. litre/1mA/h Scaled display
nor_to Scaled display Totalizer
Normaldisplay displaywith*-button
Counting time in seconds, e.g. 500 litres / mA /min, set time to 60 s (=1 min).
Displayvaluewherecountingstarts(batching).Defaultvalue=0.
WhencountingdownwardsdisplaywillshownegativevaluesiftheStartvalueissettozero.
Haltvaluepreventscountingunderthisdisplayvalue(posive).By givinganynegative
value,totalizer canbecounted also downwhen input isunder Low -value(eg. < 4mA).
Forexample: Scale4-20mA=0-1000, Halt=10(4.16mA), countingonlybetween10....1000.
Undo Returnwithoutresettingthetotalizervalue(counter)
rESEt TotalizedvalueresettoStart-valuesettinga(newbatch)
Press and hold ➤➤
➤➤
➤-button for 2 seconds. See also remote reset, page 10.
Digital filter for input signal (damping) selectable 0,01...1.0 (1.0= no filter)
0.2 = factory setting = 2 new and 8 old values
1= display update time 15/s. 7= 2/s. (average of 7 measurements)
Min/maxfunction:On=enabled,Off=disabled
IndicatorsM1=min,M2=max
For factory settings
Secret code settings ▼(▲back): ALCod= Code for alarm settings
CFCode= Code for configuration. See also section 'Secret code setting.
You can enter configuration stage by pressing two seconds ▲▲
▲▲
▲and ★★
★★
★keys at
sametime.Byarrowkeysyoucanmoveupwardsanddownwardsinmainmenu.
By pressing ➤➤
➤➤
➤key you can enter configuration stage at wished point. From
setting stageyou can skipdirect to savestage or toprevious levelwith ★★
★★
★-key.