August 2009
1638 Dew Point Analyser 5
SPECIFICATIONS
Inputs
•Zirconia oxygen probe, heated or unheated
•Furnace thermocouple, field selectable as type K or R
•Process purge complete / main flame established safety interlock (for heated probes)
•Purge pressure switch
•Remote alarm accept
Outputs
•Two linearised 4 to 20 mA DC outputs, max. load 1000Ω
•Common alarm relay
•Three other alarm relays with selectable functions
Computer
•RS 232-C or RS 485 for connection of a computer terminal or printer for diagnostics of the analyser, sensing
probe, or process. This connection is suitable for network connection to computers, DCSs or PLCs using
MODBUS protocol.
Range of Output 1
Field selectable from the following:
Output Zero Range Span Range
Dew point -60°C to +20°C -40°C to +40°C
-76°F to +84°F -56°F to +104°F
Linear oxygen 0 to 99 % 1 to 100 %
Oxygen, reducing range 1.0 -16 % oxygen 1.0-30 % oxygen
If 2 probes are used -
Linear oxygen, probe 1 0 to 99 % 1 to 100 %
Reducing oxygen output probe 1 1.0 -16 % oxygen 1.0-30 % oxygen
Dew point, probe 1 -60°C to +20°C -40°C to +40°C
-76°F to +84°F -56°F to +104°F
Linear O2, probe 1 & 2 averaged 0 to 99 % 1 to 100 %
Reducing O2, probe 1 & 2 averaged 1.0 -16 % oxygen 1.0-30 % oxygen
Dew point, probe 1 & 2 averaged -60°C to +20°C -40°C to +40°C
-76°F to +84°F -56°F to +104°F
Note:
The average dew point, linear O2and Reducing O2selections will automatically select to the operating probe if one
probe fails, or selects probe 1 if both probes fail.
Range of Output 2
Field selectable from the following:
Output Zero Range Span Range
Linear O20 to 99 % 1 to 100 %
Oxygen, reducing range 1.0 -16 % oxygen 1.0-30 % oxygen
Dew point -60°C to +20°C -40°C to +40°C
-76°F to +68°F -40°F to +104°F
Sensor EMF 0 to 1200 mV in 100 mV steps 100 to 1300 mV in 100 mV steps
Aux Temperature 0 to 1300°C in 100°C steps 100 to 1400°C in 100°C steps
If 2 probes are used -
Linear O2, probe 2 0 to 99 % 1 to 100 %
Oxygen, reducing range, probe 2 1.0 -16 % oxygen 1.0-30 % oxygen
Dew point, probe 2 -60°C to +20°C -40°C to +40°C
-76°F to +68°F -40°F to +104°F
Probe EMF 1 0 to 1200 mV in 100 mV steps 100 to 1300 mV in 100 mV steps
Probe EMF 2 0 to 1200 mV in 100 mV steps 100 to 1300 mV in 100 mV steps