1. Introduction
2. Measurement Reading Screens
3.
4. Installation
Programming
3.1 Setup Screens
3.1.1 Meter Type Selection
3.1.2 Normal Meter
3.1.2.1 Input start value
3.1.2.2 Display end value
3.1.2.3 Display start value
3.1.2.4 Bent function selection
3.1.2.5 Input bent value
3.1.2.6 Display bent value
3.1.3 Power Factor Meter
3.1.3.1 Input start value
3.1.3.2 PF end value
3.1.3.3 PF start value
4.1 EMC Installation Requirements
4.2 Case Dimensions and Panel Cut-out
4.3 Wiring
4.4 Auxiliary Supply
4.5 Fusing
4.6 Earth / Ground Connections
5. Connection Diagrams
Specifications
6.
DC Voltmeter : Eine Voltage
+
DC Ammeter : Eine Current
+
3.2 Other Display Indications
3

mV mA
Available Models :
1. Introduction
The Eine+ is a series of panel mounted 96 x 96mm and
48x96 Digital Panel Meters for the measurement of
DC Voltage and Current.
The instrument integrates accurate measurement technology.
The parameters are displayed with Ultra high bright LED
display with 14mm Digit height. Programmable DPM can be
configured and programmed on site for the following:
Input start and bent values, Display start, end and bent values.
The front panel has two push buttons through which the user
can enter into programming mode and can configure the
meter.
1. Eine+ mV 2. Eine+ mA 3. Eine+ V 4. Eine+ A
VA
A
UP DOWN
4

2. Measurement Reading Screen
In normal operation the user is presented with the measurement reading screens.
These screens may be scrolled through keys one at a time in incremental order by
pressing the key and in decrementing order by pressing key.
Screen 1 : Voltage
Screen 3 : Ambient Temperature (°F)
Screen 1 : Current
Screens of Eine
+
mA mV
Screen 2 : Ambient Temperature (°C)
mV mV
A flashing message of “°F” will be shown
at intervals of 1.5 sec.
Note: All the screen shown in manual is for reference(Screens are applicable for
both models i.e., 48x96 & 96x96)
5

A
Exit from setup Parameter to Main Display
Press both keys to enter Setup
PF
(PF display)
(Sec 3.1.3)
inP.S
(Input start)
(Sec 3.1.3.1)
nor
(normal display)
(Sec 3.1.2)
Eine+ Setup Parameter Screen
tYPE
(Meter Type)
(Sec 3.1.1)
(PF end value)
PF-E
(Sec 3.1.3.2)
PF-S
(PF start value)
(Sec 3.1.3.3)
inP.S
(Input start)
(Sec 3.1.2.1)
dSP.E
(Display end value)
(Sec 3.1.2.2)
dSP.S
(Display start value)
(Sec 3.1.2.3)
bEnt
(Bent function on/off)
(Sec 3.1.2.4)
inP.b
(Input bent value)
(Sec 3.1.2.5)
dSP.b
(Display bent value)
(Sec 3.1.2.6)
A
A
OFF
ON
6

3. Programming
The following sections comprise step by step procedures for configuring the meter
for individual user requirements. To access the set-up screens press and hold the
“ ” and “ ” keys simultaneously.
This will take the User into the Meter Type Screen. (section 3.1.1)
3.1 Setup Screens
3.1.1 Meter type
This screen is used to set the meter type. The user can
scroll between the following options using the “ up” key
nor: normal operation
PF: operate as Power Factor DPM
Pressing the “ down” key will accept the present type
and advance to Input Start value selection(section 3.1.2.1).
Pressing the “ up” key again will enter the meter type
edit mode and user can scroll through the types available.
Pressing the “ down” key advances to the meter type
confirmation menu.
Meter Type Confirmation
This screen will only appear following the edit of meter type.
Pressing the “ down” key sets the displayed type and
will advance to Input Start value selection (section 3.1.2.1).
7

This screen allows the user to set the Start value of Input.
The Start value can be set from -100% to 20% of Input End value for all models,
except for 500 & 1000V models in which Start value range is from 0 to 20% of Input
End value, where Input End value is the Nominal Input value of the meter.
3.1.2.1 Input Start Value (inP.S)
3.1.2 Normal Meter
NOTE: Decimal point will be fixed depending upon the Nominal Input value of meter.
Eg. For 20mA model, decimal pt. will be fixed after second digit. Hence, Input start
value of 04.00 will correspond to 4mA.
Initially, ‘inP.S’ message is displayed on screen following
which user can start editing Input Start value by pressing
the “ ” key.
After entering the edit mode, by pressing the “ ”key user
can turn the ‘-’ symbol On or Off. If ‘-’ symbol is On then
Input Start value will be set as negative.
Pressing the “ ” key will set the ‘-’ symbol and editing of
first digit will start. By pressing the “ ”key user can scroll
the value from 0 to 9 or it will get restricted by Input End
value.
Pressing the “ ” key will set the first digit and advance the operation to next digit.
Similarly, user can edit the remaining three digits.
After all four digits have been set, pressing the “ ” key will advance to the Input
Start value confirmation mode.
Pressing the “ ” key will re-enter into the Start value edit mode.
Pressing the “ ” key will confirm the set value. If meter type had been earlier
selected as ‘nor’ then operation will advance to Display End value selection
(section 3.1.2.2) , otherwise if ‘PF’ had been selected then it will advance to PF end
value selection (section 3.1.3.2).
8

3.1.2.2 Display End Value (dSP.E)
This screen allows the user to set the Display End value which will correspond to the
Input End value.
The End value can be set from -9999 to -10 or +10 to +9999 counts.
Initially,‘dSP.E’ message is displayed on screen following which user can start editing
Display End value by pressing the “ ” key.
Instead on pressing the “ ” key will confirm the present value and advance to
Display Start value selection (section 3.1.2.3).
After entering the edit mode, by pressing the “ ” key
user can turn the ‘-’ symbol On or Off. If ‘-’ symbol is On
then Display End value will be set as negative else it will
be positive.
Pressing the “ ” key once will set the ‘-’ symbol.
Next, on pressing “ ” key editing of Decimal point will
start. With each press of the same key, the Decimal point
will shift towards right by one digit.
Pressing the “ ” key will set the decimal point position
and advance to digit editing.
By pressing the “ ” key user can scroll the value from 0 to 9.
Pressing the “ ” key will set the first digit and advance the operation to next digit.
Similarly, user can edit the remaining three digits.
After all four digits have been set, pressing the “ ” key will advance to the Display
End value confirmation mode.
Pressing the “ ” key will re-enter into the End value edit mode.
Pressing the “ ” key will confirm the set value and advance to Display Start value
selection (section 3.1.2.3).
NOTE: The decimal point position can only be set in the Display End value screen
and the same will reflect in the measurement screen.
9

3.1.2.3 Display Start Value (dSP.S)
This screen allows the user to set the Display Start value which will correspond to the
Input Start value.
If the Display End value is negative then Start value can be set from 90% of End
value to +9999 counts, otherwise if the End value is positive then Start value can be
set from -9999 to 90% of End value.
Initially, ‘dSP.S’ message is displayed on screen following which user can start editing
Display Start value by pressing the “ ” key.
Instead on pressing the “ ” key will confirm the present value and advance to
Bent Function Selection menu (section 3.1.2.4).
After entering the edit mode, by pressing the “ ” key
user can turn the ‘-’ symbol On or Off. If ‘-’ symbol is On
then Display Start value will be set as negative else it will
be positive.
Pressing the “ ” key will set the ‘-’ symbol and editing
of first digit will start.
By pressing the “ ” key user can scroll the value from
0 to 9.
Pressing the “ ” key will set the first digit and advance the operation to next digit.
Similarly, user can edit the remaining three digits.
After all four digits have been set, pressing the “ ” key will advance to the Display
Start value confirmation mode.
Pressing the “ ” key will re-enter into the Start value edit mode.
Pressing the “ ” key will confirm the set value and advance to Bent Function
Selection menu (section 3.1.2.4).
10

3.1.2.4 Bent Function Selection
This screen allows the user to enable or disable the bent characteristics function.
This is applicable only to Normal meter type, and not to PF meter.
Initially, ‘bEnt’ message is displayed on screen.
On pressing the “ ” key, the meter will enter the Bent
function edit mode and user can scroll between yes and no.
YES: Bent function is enabled
no: Bent function is disabled
Pressing the “ ” key will accept the displayed condition.
If Bent function is enabled, then the meter will advance to
Input Bent value selection (section 3.1.2.5) otherwise will
exit the Setup menu and return to measurement mode.
3.1.2.5 Input Bent Value (inP.b)
This screen appears only when Bent Function has been enabled. This screen allows
the user to set the Input Bent value. The Bent value can be set in-between
(Input Start value + (1.5% of Input End value)) and 98.5% of Input End value.
Initially, ‘inP.b’ message is displayed on screen following
which user can start editing Input Bent value by pressing
the “ ” key.
Instead on pressing the “ ” key will confirm the present
value and advance to Display Bent value selection
(section 3.1.2.6).
After entering the edit mode, by pressing the “ ” key user
can turn the ‘-’ symbol On or Off. If ‘-’ symbol is On then
Input Bent value will be set as negative.
Pressing the “ ” key will set the ‘-’ symbol and editing of
first digit will start. By pressing the “ ” key user can scroll
the value from 0 to 9 or it will get restricted depending upon