Shenzhen Linshang Technology Co., Ltd.
2019-7-25 Page 4 of 9
A. Trigger mode: (the measurement mode of Auto and Manual)
Short press “▲” or “▼”to choose Auto/Manual
Choose “Auto”-the auto trigger mode, the meter will start and stop automatically.
Choose “Manual”-the measurement will start and stop manually.
Short press the “ENTER” key and enter into the next setting interface. If Auto mode is selected, next
setting interface is trigger threshold value. If manual mode is selected, next setting interface is the smooth
process.
Note: For the recording time is only 32 minutes, if the production line is very long and need a long time to
reach the UV Lamp position, the “AUTO” mode must be selected.
B. Trigger threshold (Trigger power: 1-50mW/cm² can be set)
Short press “▲” or “▼” to set the trigger threshold.
Short press the “ENTER” button and enter into the smooth process.
C. Smooth Processing (Smooth: OFF/50HZ/60HZ)
If the UV lamp is powered with alternating current, the frequency of alternating current will influence the
accuracy of the irradiance measurement. Short press “▲” or “▼”to select the power supply frequency of
the light source.
OFF: this option can be selected if no smooth processing will be performed and the UV lamp is
powered with direct current
50HZ: this option must be selected if frequency of the alternating current is 50HZ
60HZ: this option must be selected if frequency of the alternating current is 60HZ
Short press the “ENTER” button to confirm the setting and the meter enters the “Standard Selection”
interface.
D. Standard Selection
Short press the “ ” button or “ ” button to select the standard.
LS-STD (Enterprise standard): The meter is calibrated according to enterprise standard.
JJG 879-2015 (Chinese standard): The meter is calibrated according to Chinese standard.
Short press the “ENTER” button to confirm the setting and the meter enters the “STOP” interface.